DIRECTIVE 98/79/EC
OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL of 27 October 1998 on in vitro
diagnostic medical devices
THE EUROPEAN PARLIAMENT
AND THE COUNCIL OF THE EUROPEAN UNION,
Having regard to the Treaty establishing the European Community, and in particular
Article 100a thereof, Having regard to the proposal from the Commission (1),
Having regard to the opinion of the Economic and Social Committee (2),
Acting in accordance with the procedure laid down in Article 189b of the Treaty
(3),
(1) Whereas measures should be adopted for the smooth operation of the internal
market; whereas the internal market is an area without internal frontiers
in which the free movement of goods, persons, services and capital is ensured;
(2) Whereas the content and scope of the laws, regulations and administrative
provisions in force in the Member States with regard to the safety, health
protection and performance, characteristics and authorization procedures for
in vitro diagnostic medical devices are different; whereas the existence of
such disparities creates barriers to trade, and whereas the need to establish
harmonized rules has been confirmed by a comparative survey of national legislations
carried out on behalf of the Commission;
(3) Whereas the harmonization of national legislation is the only means of
removing such barriers to free trade and of preventing new barriers from arising;
whereas this objective cannot be achieved in a satisfactory manner by other
means by the individual Member States; whereas this Directive lays down only
such requirements as are necessary and sufficient to ensure, under the best
safety conditions, free movement of the in vitro diagnostic medical devices
to which it applies;
(4) Whereas the harmonized provisions must be distinguished from measures
adopted by the Member States to manage the funding of public health and sickness
insurance schemes relating directly or indirectly to such devices; whereas,
therefore, the harmonized provisions do not affect the ability of the Member
States to implement such measures provided that they comply with Community
law;
(5) Whereas in vitro diagnostic medical devices should provide patients, users
and third parties with a high level of health protection and attain the performance
levels originally attributed to them by the manufacturer; whereas, therefore,
maintenance or improvement of the level of health protection attained in the
Member States is one of the main objectives of this Directive;
(6) Whereas, in accordance with the principles set out in the Council resolution
of 7 May 1985 on a new approach to technical harmonization and standards (4),
rules regarding the design, manufacture and packaging of relevant products
must be confined to the provisions required to meet the essential requirements;
whereas, because they are essential, such requirements should replace the
corresponding national provisions; whereas the essential requirements, including
requirements to minimize and reduce risks, should be applied with discretion,
taking into account the technology and practice at the time of design and
technical and economic considerations compatible with a high level of protection
of health and safety;
(7) Whereas the major part of medical devices are covered by Council Directive
90/385/EEC of 20 June 1990 on the approximation of laws relating to active
implantable medical devices (5) and Council Directive 93/42/EEC of 14 June
1993 concerning medical devices
(6) with the exclusion of in vitro diagnostic medical devices; whereas this
Directive seeks to extend the harmonization to in vitro diagnostic medical
devices and whereas, in the interest of uniform Community rules, this Directive
is based largely on the provisions of the said two Directives;
(8) Whereas instruments, apparatus, appliances, materials or other articles,
including software, which are intended to be used for research purposes, without
any medical objective, are not regarded as devices for performance evaluation;
(9) Whereas, although internationally certified reference materials and materials
used for external quality assessment schemes are not covered by this Directive,
calibrators and control materials needed by the user to establish or verify
performances of devices are in vitro diagnostic medical devices;
(10) Whereas, having regard to the principle of subsidiarity, reagents which
are produced within health-institution laboratories for use in that environment
and are not subject to commercial transactions are not covered by this Directive;
(11) Whereas, however, devices that are manufactured and intended to be used
in a professional and commercial context for purposes of medical analysis
without being marketed are subject to this Directive;
(12) Whereas mechanical laboratory equipment especially designed for in vitro
diagnostic examinations falls within the scope of this Directive and whereas,
therefore, in order to harmonize the relevant directives, Directive 98/37/EC
of the European Parliament and of the Council of 22 June 1998 on the approximation
of the laws of the Member States relating to machinery (7), should be appropriately
amended to bring it into line with this Directive;
(13) Whereas this Directive should include requirements regarding the design
and manufacture of devices emitting ionizing radiation; whereas this Directive
does not affect the application of Council Directive 96/29/Euratom of 13 May
1996 laying down basic safety standards for the protection of the health of
workers and the general public against the dangers arising from ionizing radiation
(8);
(14) Whereas, since electromagnetic compatibility aspects form an integral
part of the essential requirements of this Directive, Council Directive 89/336/EEC
of 2 May 1989 on the approximation of the laws of the Member States relating
to electromagnetic compatibility (9) does not apply;
(15) Whereas, in order to ease the task of proving conformity with the essential
requirements and to enable conformity to be verified, it is desirable to have
harmonized standards in respect of the prevention of risks associated with
the design, manufacture and packaging of medical devices; whereas such harmonized
standards are drawn up by private-law bodies and should retain their status
as non-mandatory texts; whereas, to this end, the European Committee for Standardization
(CEN) and the European Committee for Electrotechnical Standardization (Cenelec)
are recognized as the competent bodies for the adoption of harmonized standards
in accordance with the general guidelines on cooperation between the Commission
and those two bodies signed on 13 November 1984;
(16) Whereas, for the purpose of this Directive, a harmonized standard is
a technical specification (European standard of harmonization document) adopted,
on a mandate from the Commission, by CEN or Cenelec or by both of those bodies
in accordance with Directive 98/34/EC of the European Parliament and of the
Council of 22 June 1998 laying down a procedure for the provision of information
in the field of technical standards and regulations (10), and pursuant to
the abovementioned general guidelines;
(17) Whereas, by way of exception to the general principles, the drawing up
of common technical specifications takes account of a current practice in
some Member States whereby for selected devices mainly used for the evaluation
of the safety of blood supply and of organ donation, such specifications are
adopted by the public authorities; whereas it is appropriate that these particular
specifications should be replaced by common technical specifications; whereas
these common technical specifications can be used for performance evaluation
and reevaluation;
(18) Whereas scientific experts from various interested parties could be involved
in the drafting of common technical specifications and in the examination
of other specific or general questions;
(19) Whereas manufacturing, as covered by this Directive, also includes the
packaging of the medical device, insofar as such packaging is related to the
safety and performance aspects of this device;
(20) Whereas certain devices have a limited life owing to the decline in their
performance over time, which is related, for example, to the deterioration
in their physical or chemical properties, including the sterility or integrity
of the packaging; whereas the manufacturer should determine and indicate the
period during which the device will perform as intended; whereas the labeling
should indicate the date until which the device or one of its components can
be used with complete safety;
(21) Whereas, in Decision 93/465/EEC of 22 July 1993 concerning the modules
for the various phases of the conformity assessment procedures and the rules
for the affixing and use of the CE conformity marking, which are intended
to be used in the technical harmonization directives (11), the Council laid
down harmonized conformity assessment procedures; whereas the details added
to these modules are justified by the nature of the verification required
for in vitro diagnostic medical devices and by the need for consistency with
Directives 90/385/EEC and 93/42/EEC;
(22) Whereas it is necessary, essentially for the purpose of the conformity
assessment procedures, to group in vitro diagnostic medical devices into two
main product classes; whereas, since the large majority of such devices do
not constitute a direct risk to patients and are used by competently trained
professionals, and the results obtained can often be confirmed by other means,
the conformity assessment procedures can be carried out, as a general rule,
under the sole responsibility of the manufacturer; whereas, taking account
of existing national regulations and of notifications received following the
procedure laid down in Directive 98/34/EC, the intervention of notified bodies
is needed only for defined devices, the correct performance of which is essential
to medical practice and the failure of which can cause a serious risk to health;
(23) Whereas, among the in vitro diagnostic medical devices for which intervention
of a notified body is required, the groups of products used in blood transfusion
and the prevention of AIDS and certain types of hepatitis require a conformity
assessment guaranteeing, with a view to their design and manufacture, an optimum
level of safety and reliability;
(24) Whereas the list of in vitro diagnostic medical devices to be subjected
to third-party conformity assessment needs updating, taking account of technological
progress and of developments in the field of health protection; whereas such
updating measures must be taken in line with procedure III(a) as laid down
in Council Decision 87/373/EEC of 13 July 1987 laying down the procedures
for the exercise of implementing powers conferred on the Commission (12);
(25) Whereas an agreement on a modus vivendi between the European Parliament,
the Council and the Commission concerning the implementing measures for acts
adopted in accordance with the procedure laid down in Article 189b of the
Treaty was reached on 20 December 1994 (13);
(26) Whereas medical devices should, as a general rule, bear the CE marking
indicating their conformity with the provisions of this Directive to enable
them to move freely within the Community and to be put into service in accordance
with their intended purpose;
(27) Whereas manufacturers will be able, when the intervention of a notified
body is required, to choose from a list of bodies published by the Commission;
whereas, although Member States do not have an obligation to designate such
notified bodies, they must ensure that bodies designated as notified bodies
comply with the assessment criteria laid down in this Directive;
(28) Whereas the director and staff of the notified bodies should not, themselves
or through an intermediary, have any interest in the establishments subject
to assessment and verification which is likely to compromise their independence;
(29) Whereas the competent authorities in charge of market surveillance should
be able, particularly in emergencies, to contact the manufacturer or his authorized
representative established in the Community, in order to take any protection
measures that should prove necessary; whereas cooperation and exchange of
information between Member States are necessary with a view to uniform application
of this Directive, in particular for the purpose of market surveillance; whereas
to that end it is necessary to establish and manage a database containing
data on manufacturers and their authorized representatives, on devices placed
on the market, on certificates issued, suspended or withdrawn, and on the
vigilance procedure; whereas a system of adverse incident reporting (vigilance
procedure) constitutes a useful tool for surveillance of the market, including
the performance of new devices; whereas information obtained from the vigilance
procedure as well as from external quality assessment schemes is useful for
decision-making on classification of devices;
(30) Whereas it is essential that manufacturers notify the competent authorities
of the placing on the market of 'new products` with regard both to the technology
used and the substances to be analyzed or other parameters; whereas this is
true in particular of high-density DNA probe devices (known as micro-chips)
used in genetic screening;
(31) Whereas, when a Member State considers that, as regards a given product
or group of products, it is necessary, in order to protect health and safety
and/or ensure compliance with the imperatives of public health, in accordance
with Article 36 of the Treaty, to prohibit or restrict their availability
or to subject it to special conditions, it may take any transitional measures
that are necessary and justified; whereas, in such cases, the Commission consults
the interested parties and the Member States and, if the national measures
are justified, adopts the necessary Community measures, in accordance with
procedure III(a) as laid down in Decision 87/373/EEC;
(32) Whereas this Directive covers in vitro diagnostic medical devices manufactured
from tissues, cells or substances of human origin; whereas it does not refer
to the other medical devices manufactured using substances of human origin;
whereas, therefore, work will have to continue in this connection in order
to produce Community legislation as soon as possible;
(33) Whereas, in view of the need to protect the integrity of the human person
during the sampling, collection and use of substances derived from the human
body, it is appropriate to apply the principles laid down in the Convention
of the Council of Europe for the protection of human rights and dignity of
the human being with regard to the application of biology and medicine; whereas,
furthermore, national regulations relating to ethics continue to apply;
(34) Whereas, in the interests of overall consistency between directives on
medical devices, some of the provisions of this Directive should be incorporated
into Directive 93/42/EEC, which needs to be amended accordingly;
(35) Whereas it is necessary to draw up as quickly as possible the legislation
which is lacking on medical devices manufactured using substances of human
origin, HAVE ADOPTED THIS DIRECTIVE:
Article 1
Scope, definitions
1. This Directive shall apply to in vitro diagnostic medical devices and their
accessories. For the purposes of this Directive, accessories shall be treated
as in vitro diagnostic medical devices in their own right. Both in vitro diagnostic
medical devices and accessories shall hereinafter be termed devices.
2. For the purposes of this Directive, the following definitions shall apply:
(a) 'medical device` means any instrument, apparatus, appliance, material
or other article, whether used alone or in combination, including the software
necessary for its proper application, intended by the manufacturer to be used
for human beings for the purpose of: - diagnosis, prevention, monitoring,
treatment or alleviation of disease, - diagnosis, monitoring, treatment, alleviation
or compensation for an injury or handicap, - investigation, replacement or
modification of the anatomy or of a physiological process, - control of conception,
and which does not achieve its principal intended action in or on the human
body by pharmacological, immunological or metabolic means, but which may be
assisted in its function by such means;
(b) 'in vitro diagnostic medical device` means any medical device which is
a reagent, reagent product, calibrator, control material, kit, instrument,
apparatus, equipment, or system, whether used alone or in combination, intended
by the manufacturer to be used in vitro for the examination of specimens,
including blood and tissue donations, derived from the human body, solely
or principally for the purpose of providing information: - concerning a physiological
or pathological state, or - concerning a congenital abnormality, or - to determine
the safety and compatibility with potential recipients, or - to monitor therapeutic
measures. Specimen receptacles are considered to be in vitro diagnostic medical
devices. 'Specimen receptacles` are those devices, whether vacuum-type or
not, specifically intended by their manufacturers for the primary containment
and preservation of specimens derived from the human body for the purpose
of in vitro diagnostic examination. Products for general laboratory use are
not in vitro diagnostic medical devices unless such products, in view of their
characteristics, are specifically intended by their manufacturer to be used
for in vitro diagnostic examination;
(c) 'accessory` means an article which, whilst not being an in vitro diagnostic
medical device, is intended specifically by its manufacturer to be used together
with a device to enable that device to be used in accordance with its intended
purpose. For the purposes of this definition, invasive sampling devices or
those which are directly applied to the human body for the purpose of obtaining
a specimen within the meaning of Directive 93/42/EEC shall not be considered
to be accessories to in vitro diagnostic medical devices;
(d) 'device for self-testing` means any device intended by the manufacturer
to be able to be used by lay persons in a home environment;
(e) 'device for performance evaluation` means any device intended by the manufacturer
to be subject to one or more performance evaluation studies in laboratories
for medical analyses or in other appropriate environments outside his own
premises;
(f) 'manufacturer` means the natural or legal person with responsibility for
the design, manufacture, packaging and labeling of a device before it is
placed on the market under his own name, regardless of whether these operations
are carried out by that person himself or on his behalf by a third party.
The obligations of this Directive to be met by manufacturers also apply to
the natural or legal person who assembles, packages, processes, fully refurbishes
and/or labels one or more ready-made products and/or assigns to them their
intended purpose as devices with a view to their being placed on the market
under his own name. This subparagraph does not apply to the person who, while
not a manufacturer within the meaning of the first subparagraph, assembles
or adapts devices already on the market to their intended purpose for an individual
patient;
(g) 'authorized representative` means any natural or legal person established
in the Community who, explicitly designated by the manufacturer, acts and
may be addressed by authorities and bodies in the Community instead of the
manufacturer with regard to the latter's obligations under this Directive;
(h) 'intended purpose` means the use for which the device is intended according
to the data supplied by the manufacturer on the labeling, in the instructions
for use and/or in promotional materials;
(i) 'placing on the market` means the first making available in return for
payment or free of charge of a device other than a device intended for performance
evaluation with a view to distribution and/or use on the Community market,
regardless of whether it is new or fully refurbished;
(j) 'putting into service` means the stage at which a device has been made
available to the final user as being ready for use on the Community market
for the first time for its intended purpose.
3. For the purposes of this Directive, calibration and control materials refer
to any substance, material or article intended by their manufacturer either
to establish measurement relationships or to verify the performance characteristics
of a device in conjunction with the intended use of that device.
4. For the purposes of this Directive, the removal, collection and use of
tissues, cells and substances of human origin shall be governed, in relation
to ethics, by the principles laid down in the Convention of the Council of
Europe for the protection of human rights and dignity of the human being with
regard to the application of biology and medicine and by any Member States
regulations on this matter.
5. This Directive shall not apply to devices manufactured and used only within
the same health institution and on the premises of their manufacture or used
on premises in the immediate vicinity without having been transferred to another
legal entity. This does not affect the right of Member State to subject such
activities to appropriate protection requirements.
6. This Directive shall not affect national laws which provide for the supply
of devices by a medical prescription.
7. This Directive is a specific directive within the meaning of Article 2(2)
of Directive 89/336/EEC, which shall cease to apply to devices which have
complied with this Directive.
Article 2
Placing on the market and putting into service Member States shall take all
necessary steps to ensure that devices may be placed on the market and/or
put into service only if they comply with the requirements laid down in this
Directive when duly supplied and properly installed, maintained and used in
accordance with their intended purpose. This involves the obligation of Member
States to monitor the security and quality of these devices. This Article
applies also to devices made available for performance evaluation.
Article 3
Essential requirements Devices must meet the essential requirements set out
in Annex I which apply to them, taking account of the intended purpose of
the devices concerned.
Article 4
Free movement
1. Member States shall not create any obstacle to the placing on the market
or the putting into service within their territory of devices bearing the
CE marking provided for in Article 16 if these devices have undergone conformity
assessment in accordance with Article 9.
2. Member States shall not create any obstacle to devices intended for performance
evaluation being made available for that purpose to the laboratories or other
institutions listed in the statement referred to in Annex VIII if they meet
the conditions laid down in Article 9(4) and Annex VIII.
3. At trade fairs, exhibitions, demonstrations, scientific or technical gatherings,
etc. Member States shall not create any obstacle to the showing of devices
which do not conform to this Directive, provided that such devices are not
used on specimens taken from the participants and that a visible sign clearly
indicates that such devices cannot be marketed or put into service until they
have been made to comply.
4. Member States may require the information to be supplied pursuant to Annex
I, part B, section 8 to be in their official language(s) when a device reaches
the final user. Provided that safe and correct use of the device is ensured,
Member States may authorize the information referred to in the first subparagraph
to be in one or more other official Community language(s). In the application
of this provision, Member States shall take into account the principle of
proportionality and, in particular:
(a) whether the information can be supplied by harmonized symbols or recognized
codes or other measures;
(b) the type of user anticipated for the device.
5. Where the devices are subject to other directives concerning other aspects
which also provide for the affixing of the CE marking, the latter shall indicate
that the devices also fulfil the provisions of the other directives. However,
should one or more of these directives allow the manufacturer, during a transitional
period, to choose which arrangements to apply, the CE marking shall indicate
that the devices fulfil the provisions only of those directives applied by
the manufacturer. In this case, the particulars of these directives, as published
in the Official Journal of the European Communities, must be given in the
documents, notices or instructions required by the directives and accompanying
such devices.
Article 5
Reference to standards
1. Member States shall presume compliance with the essential requirements
referred to in Article 3 in respect of devices which are in conformity with
the relevant national standards transposing the harmonized standards the reference
numbers of which have been published in the Official Journal of the European
Communities; Member States shall publish the reference numbers of such national
standards.
2. If a Member State or the Commission considers that the harmonized standards
do not entirely meet the essential requirements referred to in Article 3,
the measures to be taken by the Member States with regard to these standards
and the publication referred to in paragraph 1 of this Article shall be adopted
by the procedure defined in Article 6(2).
3. Member States shall presume compliance with the essential requirements
referred to in Article 3 in respect of devices designed and manufactured in
conformity with common technical specifications drawn up for the devices in
List A of Annex II and, where necessary, the devices in List B of Annex II.
These specifications shall establish appropriate performance evaluation and
re-evaluation criteria, batch release criteria, reference methods and reference
materials. The common technical specifications shall be adopted in accordance
with the procedure mentioned in Article 7(2) and be published in the Official
Journal of the European Communities. Manufacturers shall as a general rule
be required to comply with the common technical specifications; if for duly
justified reasons manufacturers do not comply with those specifications they
must adopt solutions of a level at least equivalent thereto. Where, in this
Directive, reference is made to harmonized standards, this is also meant to
refer to the common technical specifications.
Article 6
Committee on Standards and Technical Regulations
1. The Commission shall be assisted by the committee set up by Article 5 of
Directive 98/34/EC.
2. The representative of the Commission shall submit to the committee a draft
of the measures to be taken. The committee shall deliver its opinion on the
draft within a time limit which the chairman may lay down according to the
urgency of the matter, if necessary by taking a vote. The opinion shall be
recorded in the minutes; in addition, each Member State shall have the right
to ask to have its position recorded in the minutes. The Commission shall
take the utmost account of the opinion delivered by the committee. It shall
inform the committee of the manner in which its opinion has been taken into
account.
Article 7
Committee on Medical Devices
1. The Commission shall be assisted by the committee set up by Article 6(2)
of Directive 90/385/EEC.
2. The representative of the Commission shall submit to the committee a draft
of the measures to be taken. The committee shall deliver its opinion on the
draft within a time limit which the chairman may lay down according to the
urgency of the matter. The opinion shall be delivered by the majority laid
down in Article 148(2) of the Treaty in the case of decisions which the Council
is required to adopt on a proposal from the Commission. The votes of the representatives
of the Member States within the committee shall be weighted in the manner
set out in that Article. The chairman shall not vote. The Commission shall
adopt the measures envisaged if they are in accordance with the opinion of
the committee. If the measures envisaged are not in accordance with the opinion
of the committee, or if no opinion is delivered, the Commission shall, without
delay, submit to the Council a proposal relating to the measures to be taken.
The Council shall act by a qualified majority. If, on the expiry of a period
of three months from the date of referral to the Council, the Council has
not acted, the proposed measures shall be adopted by the Commission.
3. The committee referred to in paragraph 1 may examine any question connected
with the implementation of this Directive.
Article 8
Safeguard clause
1. Where a Member State ascertains that the devices referred to in Article
4(1), when correctly installed, maintained and used for their intended purpose
may compromise the health and/or safety of patients, users or, where applicable,
other persons, or the safety of property, it shall take all appropriate interim
measures to withdraw such devices from the market or prohibit or restrict
their being placed on the market or put into service. The Member State shall
immediately inform the Commission of any such measures, indicating the reasons
for its decision and, in particular, whether non-compliance with this Directive
is due to:
(a) failure to meet the essential requirements referred to in Article 3;
(b) incorrect application of the standards referred to in Article 5, insofar
as it is claimed that the standards have been applied;
(c) shortcomings in the standards themselves.
2. The Commission shall enter into consultation with the parties concerned
as soon as possible. Where, after such consultation, the Commission finds
that: - the measures are justified, it shall immediately so inform the Member
State which took the initiative and the other Member States; where the decision
referred to in paragraph 1 is attributed to shortcomings in the standards,
the Commission shall, after consulting the parties concerned, bring the matter
before the committee referred to in Article 6(1) within two months if the
Member State which has taken the decision intends to maintain it and shall
initiate the procedures referred to in Article 6; where the measure referred
to in paragraph 1 is attributed to problems related to the contents or to
the application of the common technical specifications, the Commission shall,
after consulting the parties concerned, bring the matter before the Committee
referred to in Article 7(1) within two months, - the measures are unjustified,
it shall immediately so inform the Member State which took the initiative
and the manufacturer or hisauthorizedd representative.
3. Where a non-complying device bears the CE marking, the competent Member
State shall take appropriate action against whomsoever affixed the marking
and shall inform the Commission and the other Member States thereof.
4. The Commission shall ensure that the Member States are kept informed of
the progress and outcome of this procedure.
Article 9
Conformity assessment procedures
1. For all devices other than those covered by Annex II and devices for performance
evaluation, the manufacturer shall, in order to affix the CE marking, follow
the procedure referred to in Annex III and draw up the EC declaration of conformity
required before placing the devices on the market. For all devices for self-testing
other than those covered by Annex II and devices for performance evaluation,
the manufacturer shall, prior to the drawing up of the aforementioned declaration
of conformity, fulfil the supplementary requirements set out in Annex III,
point 6. Instead of applying this procedure, the manufacturer may follow the
procedure referred to in paragraphs 2 or 3.
2. For all devices referred to in List A in Annex II other than those intended
for performance evaluation, the manufacturer shall, in order to affix the
CE marking either:
(a) follow the procedure relating to the EC declaration of conformity set
out in Annex IV (full quality assurance), or
(b) follow the procedure relating to EC type-examination set out in Annex
V coupled with the procedure relating to the EC declaration of conformity
set out in Annex VII (production quality assurance).
3. For all devices referred to in List B in Annex II other than those intended
for performance evaluation, the manufacturer shall for the purposes of affixing
the CE marking, follow either:
(a) the procedure relating to the EC declaration of conformity set out in
Annex IV (full quality assurance) or
(b) the procedure relating to EC type-examination set out in Annex V couplet
with: (i) the procedure relating to EC verification set out in Annex VI, or
(ii) the procedure relating to the EC declaration of conformity set out in
Annex VII (production quality assurance).
4. In the case of devices for performance evaluation, the manufacturer shall
follow the procedure referred to in Annex VIII and draw up the statement set
out in that Annex before such devices are made available. This provision does
not affect national regulations relating to the ethical aspects of carrying
out performance evaluation studies using tissues or substances of human origin.
5. During the conformity assessment procedure for a device, the manufacturer
and, if involved, the notified body shall take account of the results of any
assessment and verification operations which, where appropriate, have been
carried out in accordance with this Directive at an intermediate state of
manufacture.
6. The manufacturer may instruct hisauthorizedd representative to initiate
the procedures provided for in Annexes III, V, VI and VIII.
7. The manufacturer must keep the declaration of conformity, the technical
documentation referred to in Annexes III to VIII, as well as the decisions,
reports and certificates, established by notified bodies, and make it available
to the national authorities for inspection purposes for a period ending five
years after the last product has been manufactured. Where the manufacturer
is not established in the Community, the obligation to make the aforementioned
documentation available on request applies to hisauthorizedd representative.
8. Where the conformity assessment procedure involves intervention of a notified
body, the manufacturer, or authorized representative, may apply to a body
of his choice within the framework of tasks for which the body has been notified.
9. The notified body may require, where duly justified, any information or
data, which is necessary for establishing and maintaining the attestation
of conformity in view of the chosen procedure.
10. Decisions taken by the notified bodies in accordance with Annexes III,
IV, and V shall be valid for a maximum of five years and may be extended on
application, made at a time agreed in the contract signed by both parties,
for further periods of up to five years.
11. The records and correspondence relating to the procedures referred to
in paragraphs 1 to 4 shall be in an official language of the Member State
in which the procedures are carried out and/or in another Community language
acceptable to the notified body.
12. By way of derogation from paragraphs 1 to 4, the competent authorities
may authorize, on duly justified request, the placing on the market and putting
into service, within the territory of the Member State concerned, of individual
devices for which the procedures referred to in paragraphs 1 to 4 have not
been carried out and the use of which is in the interest of protection of
health.
13. The provisions of this Article shall apply accordingly to any natural
or legal person who manufacturers devices covered by this Directive and, without
placing them on the market, puts them into service and uses them in the context
of his professional activity.
Article 10
Registration of manufacturers and devices
1. Any manufacturer who places devices on the market under his own name shall
notify the competent authorities of the Member State in which he has his registered
place of business:
- of the address of the registered place of business,
- of information relating to the reagents, reagent products and calibration
and control materials in terms of common technological characteristics and/or
analytes of any significant change thereto including discontinuation of
placing on the market; for other devices, the appropriate indications,
- in the case of devices covered by Annex II and of devices for self-testing,
of all data allowing for identification of such devices, the analyte and,
where appropriate, diagnostic parameters as referred to in Annex I, part A,
section 3, the outcome of performance evaluation pursuant to Annex VIII, certificates
and any significant change thereto, including discontinuation of placing on
the market.
2. For devices covered by Annex II and for devices for self-testing, Member
States may request to be informed of the data allowing identification together
with the label and the instructions for use when such devices are placed on
the market and/or put into service within their territory. These measures
cannot constitute a precondition for the placing on the market and/or putting
into service of devices which are in conformity with this Directive.
3. Where a manufacturer who places devices on the market under his own name
does not have a registered place of business in a Member State, he shall designate
an authorized representative. The authorized representative shall notify the
competent authorities of the Member State in which he has his registered place
of business of all particulars as referred to in paragraph 1.
4. The notification referred to in paragraph 1 shall also include any new
device. In addition, where, in the context of such notification, a device
notified, bearing the CE marking, is a 'new product`, the manufacturer shall
indicate this fact on his notification. For the purposes of this Article,
a device is 'new` if:
(a) there has been no such device continuously available on the Community
market during the previous three years for the relevant analyte or other parameter;
(b) the procedure involves analytical technology not continuously used in
connection with a given analyte or other parameter on the Community market
during the previous three years.
5. Member States shall take all necessary measures to ensure that the notifications
referred to in paragraphs 1 and 3 are registered immediately in the databank
described in Article 12. The procedures for implementing this Article and
in particular those referring to the notification and the concept of significant
change shall be adopted in accordance with the procedure referred to in Article
7.
6. Transitionally, pending the establishment of a European databank accessible
to the competent authorities of the Member States and containing the data
relating to all devices available on the territory of the Community, the manufacturer
shall give such notification to the competent authorities of each Member State
concerned by the placing on the market. Article 11 Vigilance procedure
1. Member States shall take the necessary steps to ensure that any information
brought to their knowledge, in accordance with the provisions of this Directive,
regarding the incidents mentioned below involving devices bearing the CE marking
is recorded and evaluated centrally:
(a) any malfunction, failure or deterioration in the characteristics and/or
performance of a device, as well as any inadequacy in the labeling or the
instructions for use which, directly or indirectly, might lead to or might
have led to the death of a patient, or user or of other persons or to a serious
deterioration in their state of health;
(b) any technical or medical reason in relation to the characteristics or
performance of a device for the reasons referred to in subparagraph (a), leading
to systematic recall of devices of the same type by the manufacturer.
2. Where a Member State requires medical practitioners, the medical institutions
or the organizerss of external quality assessment schemes to inform the competent
authorities of any incidents referred to in paragraph 1, it shall take the
necessary steps to ensure that the manufacturer of the device concerned, or
his authorized representative, is also informed of the incident.
3. After carrying out an assessment, if possible together with the manufacturer,
Member States shall, without prejudice to Article 8, immediately inform the
Commission and the other Member States of the incidents referred to in paragraph
1 for which appropriate measures, including possible withdrawal, have been
taken or are contemplated.
4. Where, in the context of notification referred to in Article 10, a device
notified, bearing the CE marking, is a 'new` product, the manufacturer shall
indicate this fact on his notification. The competent authority so notified
may at any time within the following two years and on justified grounds, require
the manufacturer to submit a report relating to the experience gained with
the device subsequent to its being placed on the market.
5. The Member States shall on request inform the other Member States of the
details referred to in paragraphs 1 to 4. The procedures implementing this
Article shall be adopted in accordance with the procedure referred to in Article
7(2).
Article 12
European databank
1. Regulatory data in accordance with this Directive shall be stored in a
European databank accessible to the competent authorities to enable them to
carry out their tasks relating to this Directive on a well-informed basis.
The databank shall contain the following:
(a) data relating to registration of manufacturers and devices in accordance
with Article 10;
(b) data relating to certificates issued, modified, supplemented, suspended,
withdrawn or refused according to the procedure as laid down in Annexes III
to VII;
(c) data obtained in accordance with the vigilance procedure as defined in
Article 11.
2. Data shall be forwarded in a standardized format.
3. The procedures implementing this Article shall be adopted in accordance
with the procedure laid down in Article 7(2).
Article 13
Particular health monitoring measures
Where a Member State considers, in relation to a given product or group of
products, that, in order to ensure protection of health and safety and/or
to ensure that public health requirements are observed pursuant to Article
36 of the Treaty, the availability of such products should be prohibited,
restricted or made subject to particular requirements, it may take any necessary
and justified transitional measures. It shall then inform the Commission and
all the other Member States, giving the reasons for its decision. The Commission
shall consult the interested parties and the Member States and, where the
national measures are justified, adopt necessary Community measures in accordance
with the procedure referred to in Article 7(2).
Article 14
Amendments to Annex II, and derogation clause
1. Where a Member State considers that:
(a) the list of devices in Annex II should be amended or extended, or
(b) the conformity of a device or category of devices should be established,
by way of derogation from the provisions of Article 9, by applying one or
more given procedures taken from amongst those referred to in Article 9, it
shall submit a duly substantiated request to the Commission and ask it to
take the necessary measures. The measures shall be adopted in accordance with
the procedure laid down in Article 7(2).
2. When a measure is to be taken in accordance with paragraph 1, due consideration
shall be given to:
(a) any relevant information available from the vigilance procedures and from
external quality assessment schemes as referred to in Article 11;
(b) the following criteria:
(i) whether total reliance has to be placed on the result obtained with a
given device, this result having a direct impact on subsequent medical action,
and
(ii) whether action taken on the basis of an incorrect result obtained using
a given device could prove to be hazardous to the patient, to a third party
or to the public, in particular as a consequence of false positive or false
negative results, and
(iii) whether the involvement of a notified body would be conducive to establishing
the conformity of the device.
3. The Commission shall inform the Member States of the measures taken and,
where appropriate, publish these measures in the Official Journal of the European
Communities.
Article 15
Notified bodies
1. Member States shall notify the Commission and other Member States of the
bodies which they have designated for carrying out the tasks pertaining to
the procedures referred to in Article 9 and the specific tasks for which the
bodies have been designated. The Commission shall assign identification numbers
to these bodies, hereinafter referred to as 'notified bodies`. The Commission
shall publish a list of the notified bodies, together with the identification
numbers it has allocated to them and the tasks for which they have been notified,
in the Official Journal of the European Communities. It shall ensure that
the list is kept up to date. Member States shall not be obliged to designate
a notified body.
2. Member States shall apply the criteria set out in Annex IX for the designation
of bodies. Bodies that meet the criteria laid down in the national standards
which transpose the relevant harmonized standards shall be presumed to meet
the relevant criteria.
3. Member States shall apply continual surveillance of notified bodies to
ensure ongoing compliance with the criteria set out in Annex IX. A Member
State that has notified a body shall withdraw or restrict that notification
if it finds that the body no longer meets the criteria referred to in Annex
IX. It shall immediately inform the other Member States and the Commission
of any withdrawal of notification or any restriction placed on it.
4. The notified body and the manufacturer, or his authorized representative
established in the Community, shall lay down, by common accord, the time limits
for completion of the assessment and verification operations referred to in
Annexes III to VII.
5. The notified body shall inform the other notified bodies and the competent
authority about all certificates suspended or withdrawn and, on request, about
certificates issued or refused. It shall also make available, on request,
all additional relevant information.
6. Where a notified body finds that pertinent requirements of this Directive
have not been met or are no longer met by the manufacturer or where a certificate
should not have been issued, it shall, taking account of the principle of
proportionality, suspend or withdraw the certificate issued or place any restrictions
on it unless compliance with such requirements is ensured by the implementation
of appropriate corrective measures by the manufacturer. In the case of suspension
or withdrawal of the certificate or of any restriction placed on it or in
cases where an intervention of the competent authority may become necessary,
the notified body shall inform its competent authority thereof. The Member
State shall inform the other Member States and the Commission.
7. The notified body shall, on request, supply all relevant information and
documents, including budgetary documents, required to enable the Member State
to verify compliance with Annex IX requirements.
Article 16
CE marking
1. Devices, other than devices for performance evaluation, considered to meet
the essential requirements referred to in Article 3 must bear the CE marking
of conformity when they are placed on the market.
2. The CE marking of conformity, as shown in Annex X, must appear in a visible,
legible and indelible form on the device, where practicable and appropriate,
and on the instructions for use. The CE marking of conformity must also appear
on the sales packaging. The CE marking shall be accompanied by the identification
number of the notified body responsible for implementation of the procedures
set out in Annexes III, IV, VI and VII.
3. It is prohibited to affix marks or inscriptions which are likely to mislead
third parties with regard to the meaning or the graphics of the CE marking.
Any other mark may be affixed to the device, to the packaging or to the instruction
leaflet accompanying the device provided that the visibility and legibility
of the CE marking is not thereby reduced.
Article 17
Wrongly affixed CE marking
1. Without prejudice to Article 8:
(a) where a Member State establishes that the CE marking has been wrongly
affixed, the manufacturer or his authorized representative shall be obliged
to end the infringement under conditions imposed by the Member State;
(b) where non-compliance continues, the Member State must take all appropriate
measures to restrict or prohibit the placing on the market of the product
in question or to ensure that it is withdrawn from the market, in accordance
with the procedure in Article 8.
2. The provisions stated in paragraph 1 shall also apply where the CE marking
has been affixed in accordance with the procedures in this Directive, but
inappropriately, on products that are not covered by this Directive.
Article 18
Decisions in respect of refusal or restriction
1. Any decision taken pursuant to this Directive:
(a) to refuse or restrict the placing on the market or any making available
or putting into service of a device, or
(b) to withdraw devices from the market, shall state the exact grounds on
which it is based. Such decisions shall be notified without delay to the party
concerned, who shall at the same time be informed of the remedies available
to him under the national law in force in the Member State in question and
of the time limits to which such remedies are subject.
2. In the event of a decision as referred to in paragraph 1, the manufacturer
or his authorized representative shall have an opportunity to put forward
his point of view in advance, unless such consultation is not possible because
of the urgency of the measure to be taken as justified in particular by public
health requirements.
Article 19
Confidentiality
Without prejudice to national law and practice on medical secrecy, Member
States shall ensure that all the parties involved in the application of this
Directive are bound to observe confidentiality with regard to information
obtained in carrying out their tasks. This does not affect the obligations
of Member States and notified bodies with regard to mutual information and
the dissemination of warnings, nor the obligations of the persons concerned
to provide information under criminal law.
Article 20
Cooperation between Member States Member States shall take appropriate measures
to ensure that competent authorities charged with the implementation of this
Directive cooperate with each other and convey to each other the information
necessary to ensure application in compliance with this Directive.
Article 21
Amendment of directives
1. In Directive 89/392/EEC, the second indent of Article 1(3), 'machinery
for medical use, used in direct contact with patients` shall be replaced by
the following: '- medical devices,`.
2. Directive 93/42/EEC is hereby amended as follows:
(a) in Article 1(2): - point (c) shall be replaced by the following: '(c)
"in vitro diagnostic medical device" means any medical device which is a reagent,
reagent product, calibrator, control material, kit, instrument, apparatus,
equipment or system, whether used alone or in combination, intended by the
manufacturer to be used in vitro for the examination of specimens, including
blood and tissue donations, derived from the human body, solely or principally
for the purpose of providing information: - concerning a physiological or
pathological state, or - concerning a congenital abnormality, or - to determine
the safety and compatibility with potential recipients, or - to monitor therapeutic
measures. Specimen receptacles are considered to be in vitro diagnostic medical
devices. "Specimen receptacles" are those devices, whether vacuum-type or
not, specifically intended by their manufacturers for the primary containment
and preservation of specimens derived from the human body for the purpose
of in vitro diagnostic examination. Products for general laboratory use are
not in vitro diagnostic medical devices unless such products, in view of their
characteristics, are specifically intended by their manufacturer to be used
for in vitro diagnostic examination;` - point (i) shall be replaced by the
following: '(i) "putting into service" means the stage at which a device has
been made available to the final user as being ready for use on the Community
market for the first time for its intended purpose;` - the following point
shall be added: '(j) "authorized representative" means any natural or legal
person established in the Community who, explicitly designated by the manufacturer,
acts and may be addressed by authorities and bodies in the Community instead
of the manufacturer with regard to the latter's obligations under this Directive;`
(b) Article 2 shall be replaced by the following: 'Article 2 Placing on the
market and putting into service Member States shall take all necessary steps
to ensure that devices may be placed on the market and/or put into service
only if they comply with the requirements laid down in this Directive when
duly supplied and properly installed, maintained and used in accordance with
their intended purpose.`
(c) the following paragraph shall be added to Article 14(1): 'For all medical
devices of classes IIb and III, Member States may request to be informed of
all data allowing for identification of such devices together with the label
and the instructions for use when such devices are put into service within
their territory.`
(d) the following Articles shall be inserted: 'Article 14a European databank
1. Regulatory data in accordance with this Directive shall be stored in a
European database accessible to the competent authorities to enable them to
carry out their tasks relating to this Directive on a well-informed basis.
The databank shall contain the following: (a) data relating to registration
of manufacturers and devices in accordance with Article 14; (b) data relating
to certificates issued, modified, supplemented, suspended, withdrawn or refused
according to the procedures, as laid down in Annexes II to VII; (c) data obtained
in accordance with the vigilance procedure as defined in Article 10; 2. Data
shall be forwarded in a standardized format.
3. The procedures implementing this Article shall be adopted in accordance
with the procedure laid down in Article 7(2). Article 14b Particular health
monitoring measures Where a Member State considers, in relation to a given
product or group of products, that, in order to ensure protection of health
and safety and/or to ensure that public health requirements are observed pursuant
to Article 36 of the Treaty, the availability of such products should be prohibited,
restricted or subjected to particular requirements, it may take any necessary
and justified transitional measures. It shall then inform the Commission and
all the other Member States giving the reasons for its decision. The Commission
shall, whenever possible, consult the interested parties and the Member States
and, where the national measures are justified, adopt necessary Community
measures in accordance with the procedure referred to in Article 7(2).` (e)
the following paragraphs shall be added to Article 16: '5. The notified body
shall inform the other notified bodies and the competent authority about all
certificates suspended or withdrawn and, on request, about certificates issued
or refused. It shall also make available, on request, all additional relevant
information.
6. Where a notified body finds that pertinent requirements of this Directive
have not been met or are no longer met by the manufacturer or where a certificate
should not have been issued, it shall, taking account of the principle of
proportionality, suspend or withdraw the certificate issued or place any restrictions
on it unless compliance with such requirements is ensured by the implementation
of appropriate corrective measures by the manufacturer. In the case of suspension
or withdrawal of the certificate or of any restriction placed on it or in
cases where an intervention of the competent authority may become necessary,
the notified body shall inform its competent authority thereof. The Member
State shall inform the other Member States and the Commission.
7. The notified body shall, on request, supply all relevant information and
documents including budgetary documents, required to enable the Member State
to verify compliance with Annex XI requirements.` (f) the following paragraph
shall be added to Article 18: 'Those provisions shall also apply where the
CE marking has been affixed in accordance with the procedures in this Directive,
but inappropriately, on products that are not covered by this Directive.`
(g) in Article 22(4), the first subparagraph shall be replaced by the following:
'4. Member States shall accept: - devices which conform to the rules in force
in their territory on 31 December 1994 being placed on the market during a
period of five years following the adoption of this Directive, and - the aforementioned
devices being put into service until 30 June 2001 at the latest.` (h) Annex
II, section 6.2, Annex III, section 7.1, Annex V, section 5.2 and Annex VI,
section 5.2 shall be deleted; (i) in Annex XI, section 3 the following sentence
shall be inserted after the second sentence: 'This presupposes the availability
of sufficient scientific staff within the organization who possess experience
and knowledge sufficient to assess the medical functionality and performance
of devices for which it has been notified, having regard to the requirements
of this Directive and, in particular, those set out in Annex I.`
Article 22
Implementation, transitional provisions
1. Member States shall adopt and publish the laws, regulations and administrative
provisions necessary to comply with this Directive not later than 7 December
1999. They shall immediately inform the Commission thereof. Member States
shall apply these provisions with effect from 7 June 2000. When Member States
adopt these provisions, they shall contain a reference to this Directive or
shall be accompanied by such a reference on the occasion of their official
publication. The methods of making such a reference shall be laid down by
the Member States.
2. Member States shall communicate to the Commission the texts of the main
provisions of domestic law which they adopt in the field governed by this
Directive.
3. The Committee referred to in Article 7 may undertake its tasks from the
date of entry into force of this Directive. The Member States may take the
measures referred to in Article 15 as from the entry force of this Directive.
4. Member States shall take the necessary action to ensure that the notified
bodies which are responsible pursuant to Article 9 for conformity assessment
take account of any relevant information regarding the characteristics and
performance of such devices, including in particular the results of any relevant
test and verification already carried out under pre-existing national law,
regulations or administrative provisions in respect of such devices.
5. During a period of five years following the entry into force of this Directive,
Member States shall accept the placing on the market of devices which conform
to the rules in force in their territory on the date on which this Directive
enters into force. For an additional period of two years, the said devices
may be put into service.
Article 23
This Directive shall enter into force on the day of its publication in the
Official Journal of the European Communities.
Article 24
This Directive is addressed to the Member States.
ANNEX I
ESSENTIAL REQUIREMENTS
A. GENERAL REQUIREMENTS
1. The devices must be designed and manufactured in such a way that, when
used under the conditions and for the purposes intended, they will not compromise,
directly or indirectly, the clinical condition or the safety of the patients,
the safety or health of users or, where applicable, other persons, or the
safety of property. Any risks which may be associated with their use must
be acceptable when weighed against the benefits to the patient and be compatible
with a high level of protection of health and safety.
2. The solutions adopted by the manufacturer for the design and construction
of the devices must conform to safety principles, taking account of the generally
acknowledged state of the art. In selecting the most appropriate solutions,
the manufacturer must apply the following principles in the following order:
- eliminate or reduce risks as far as possible (inherently safe design and
construction), - where appropriate take adequate protection measures in relation
to risks that cannot be eliminated, - inform users of the residual risks due
to any shortcomings of the protection measures adopted.
3. The devices must be designed and manufactured in such a way that they are
suitable for the purposes referred to in Article 1(2)(b), as specified by
the manufacturer, taking account of the generally acknowledged state of the
art. They must achieve the performances, in particular, where appropriate,
in terms of analytical sensitivity, diagnostic sensitivity, analytical specificity,
diagnostic specificity, accuracy, repeatability, reproducibility, including
control of known relevant interference, and limits of detection, stated by
the manufacturer. The traceability of values assigned to calibrators and/or
control materials must be assured through available reference measurement
procedures and/or available reference materials of a higher order.
4. The characteristics and performances referred to in sections 1 and 3 must
not be adversely affected to such a degree that the health or the safety of
the patient or the user and, where applicable, of other persons, are compromised
during the lifetime of the device as indicated by the manufacturer, when the
device is subjected to the stresses which can occur during normal conditions
of use. When no lifetime is stated, the same applies for the lifetime reasonably
to be expected of a device of that kind, having regard to the intended purpose
and the anticipated use of the device.
5. The devices must be designed, manufactured and packed in such a way that
their characteristics and performances during their intended use will not
be adversely affected under storage and transport conditions (temperature,
humidity, etc.) taking account of the instructions and information provided
by the manufacturer.
B. DESIGN AND MANUFACTURING REQUIREMENTS
1. Chemical and physical properties
1.1. The devices must be designed and manufactured in such a way as to achieve
the characteristics and performances referred to in section A on the 'General
requirements`. Particular attention must be paid to the possibility of impairment
of analytical performance due to incompatibility between the materials used
and the specimens (such as biological tissues, cells, body fluids and micro-organisms)
intended to be used with the device, taking account of its intended purpose.
1.2. The devices must be designed, manufactured and packed in such a way as
to reduce as far as possible the risk posed by product leakage, contaminants
and residues to the persons involved in the transport, storage and use of
the devices, taking account of the intended purpose of the products.
2. Infection and microbial contamination
2.1. The devices and their manufacturing processes must be designed in such
a way as to eliminate or reduce as far as possible the risk of infection to
the user or other persons. The design must allow easy handling and, where
necessary, reduce as far as possible contamination of, and leakage from, the
device during use and, in the case of specimen receptacles, the risk of contamination
of the specimen. The manufacturing processes must be appropriate for these
purposes.
2.2. Where a device incorporates biological substances, the risks of infection
must be reduced as far as possible by selecting appropriate donors and appropriate
substances and by using appropriate, validated inactivation, conservation,
test and control procedures.
2.3. Devices labeled either as 'STERILE` or as having a special microbiological
state must be designed, manufactured and packed in an appropriate pack, according
to procedures suitable for ensuring that they remain in the appropriate microbiological
state indicated on the label when placed on the market, under the storage
and transport conditions specified by the manufacturer, until the protective
packaging is damaged or opened.
2.4. Devices labeled either as 'STERILE` or as having a special microbiological
state must have been processed by an appropriate, validated method.
2.5. Packaging systems for devices other than those referred to in section
2.3 must keep the product without deterioration at the level of cleanliness
indicated by the manufacturer and, if the devices are to be sterilized prior
to use, reduce as far as possible the risk of microbial contamination. Steps
must be taken to reduce as far as possible microbial contamination during
selection and handling of raw materials, manufacture, storage and distribution
where the performance of the device can be adversely affected by such contamination.
2.6. Devices intended to be sterilized must be manufactured in appropriately
controlled (e.g. environmental) conditions.
2.7. Packaging systems for non-sterile devices must keep the product without
deterioration at the level of cleanliness stipulated and, if the devices are
to be sterilized prior to use, minimize the risk of microbial contamination;
the packaging system must be suitable taking account of the method of sterilization
indicated by the manufacturer.
3. Manufacturing and environmental properties
3.1. If the device is intended for use in combination with other devices or
equipment, the whole combination, including the connection system, must be
safe and must not impair the specified performances of the devices. Any restrictions
on use must be indicated on the label and/or in the instructions for use.
3.2. Devices must be designed and manufactured in such a way as to reduce
as far as possible the risks linked to their use in conjunction with materials,
substances and gases with which they may come into contact during normal conditions
of use.
3.3. Devices must be designed and manufactured in such a way as to remove
or reduce as far as possible: - the risk of injury linked to their physical
features (in particular aspects of volume x pressure, dimension and, where
appropriate, ergonomic features), - risks linked to reasonably foreseeable
external influences, such as magnetic fields, external electrical effects,
electrostatic discharge, pressure, humidity, temperature or variations in
pressure or acceleration or accidental penetration of substances into the
device. Devices must be designed and manufactured in such a way as to provide
an adequate level of intrinsic immunity of electromagnetic disturbance to
enable them to operate as intended.
3.4. Devices must be designed and manufactured in such a way as to reduce
as far as possible the risks of fire or explosion during normal use and in
single fault condition. Particular attention must be paid to devices whose
intended use includes exposure to or use in association with flammable substances
or substances which could cause combustion.
3.5. Devices must be designed and manufactured in such a way as to facilitate
the management of safe waste disposal.
3.6. The measuring, monitoring or display scale (including colour change and
other visual indicators) must be designed and manufactured in line with ergonomic
principles, taking account of the intended purpose of the device.
4. Devices which are instruments or apparatus with a measuring function
4.1. Devices which are instruments or apparatus having a primary analytical
measuring function must be designed and manufactured in such a way as to provide
adequate stability and accuracy of measurement within appropriate accuracy
limits, taking into account the intended purpose of the device and of available
and appropriate reference measurement procedures and materials. The accuracy
limits have to be specified by the manufacturer.
4.2. When values are expressed numerically, they must be given in legal units
conforming to the provisions of Council Directive 80/181/EEC of 20 December
1979 on the approximation of the laws of the Member States relating to units
of measurement (1).
5. Protection against radiation
5.1. Devices shall be designed, manufactured and packaged in such a way that
exposure of users and other persons to the emitted radiation is minimized.
5.2. When devices are intended to emit potentially hazardous, visible and/or
invisible radiation, they must as far as possible be: - designed and manufactured
in such a way as to ensure that the characteristics and the quantity of radiation
emitted can be controlled and/or adjusted, - fitted with visual displays and/or
audible warnings of such emissions.
5.3. The operating instructions for devices emitting radiation must give detailed
information as to the nature of the emitted radiation, means of protecting
the user, and on ways of avoiding misuse and of eliminating the risks inherent
in installation.
6. Requirements for medical devices connected to or equipped with an energy
source
6.1. Devices incorporating electronic programmable systems, including software,
must be designed to ensure the repeatability, reliability and performance
of these systems according to the intended use.
6.2. Devices must be designed and manufactured in such a way as to minimize
the risks of creating electromagnetic perturbation which could impair the
operation of other devices or equipment in the usual environment.
6.3. Devices must be designed and manufactured in such a way as to avoid,
as far as possible, the risk of accidental electric shocks during normal use
and in single fault condition, provided the devices are installed and maintained
correctly.
6.4. Protection against mechanical and thermal risks
6.4.1. Devices must be designed and manufactured in such a way as to protect
the user against mechanical risks. Devices must be sufficiently stable under
the foreseen operating conditions. They must be suitable to withstand stresses
inherent in the foreseen working environment, and to retain this resistance
during the expected life of the devices, subject to any inspection and maintenance
requirements as indicated by the manufacturer. Where there are risks due to
the presence of moving parts, risks due to break-up or detachment, or leakage
of substances, then appropriate protection means must be incorporated. Any
guards or other means included with the device to provide protection, in particular
against moving parts, must be secure and must not interfere with access for
the normal operation of the device, or restrict routine maintenance of the
device as intended by the manufacturer.
6.4.2. Devices must be designed and manufactured in such a way as to reduce
to the lowest possible level the risks arising from vibration generated by
the devices, taking account of technical progress and of the means available
for limiting vibrations, particularly at source, unless the vibrations are
part of the specified performance.
6.4.3. Devices must be designed and manufactured in such a way as to reduce
as far as possible the risks arising from the noise emitted, taking account
of technical progress and of the means available to reduce noise, particularly
at source, unless the noise emitted is part of the specified performance.
6.4.4. Terminals and connectors to electricity, gas or hydraulic and pneumatic
energy supplies which the user has to handle must be designed and manufactured
in such a way as to minimize all possible risks.
6.4.5. Accessible parts of the devices (excluding the parts of areas intended
to supply heat or reach given temperatures) and their surroundings must not
attain potentially dangerous temperatures under normal use.
7. Requirements for devices for self-testing Devices for self-testing must
be designed and manufactured in such a way that they perform appropriately
for their intended purpose taking into account the skills and the means available
to users and the influence resulting from variation that can reasonably be
anticipated in users' technique and environment. The information and instructions
provided by the manufacturer should be easily understood and applied by the
user.
7.1. Devices for self-testing must be designed and manufactured in such a
way as to: - ensure that the device is easy to use by the intended lay user
at all stages of the procedure, and - reduce as far as practicable the risk
of user error in the handling of the device and in the interpretation of the
results.
7.2. Devices for self-testing must, where reasonably possible, include user
control, i.e. a procedure by which the user can verify that, at the time of
use, the product will perform as intended.
8. Information supplied by the manufacturer
8.1. Each device must be accompanied by the information needed to use it safely
and properly, taking account of the training and knowledge of the potential
users, and to identify the manufacturer. This information comprises the data
on the label and in the instructions for use. As far as practicable and appropriate,
the information needed to use the device safely and properly must be set out
on the device itself and/or, where appropriate, on the sales packaging. If
individual full labeling of each unit is not practicable, the information
must be set out on the packaging and/or in the instructions for use supplied
with one or more devices. Instructions for use must accompany or be included
in the packaging of one or more devices. In duly justified and exceptional
cases no such instructions for use are needed for a device if it can be used
properly and safely without them.
8.2. Where appropriate, the information to be supplied should take the form
of symbols. Any symbol and identification colour used must conform to the
harmonized standards. In areas for which no standards exist, the symbols and
colour used must be described in the documentation supplied with the device.
8.3. In the case of devices containing or a preparation which may be considered
as being dangerous, taking account of the nature and quantity of its constituents
and the form under which they are present, relevant danger symbols and labeling
requirements of Directive 67/548/EEC (2) and Directive 88/379/EEC (3) shall
apply. Where there is insufficient space to put all the information on the
device itself or on its label, the relevant danger symbols shall be put on
the label and the other information required by those Directives shall be
given in the instructions for use. The provisions of the aforementioned Directives
on the safety data sheet shall apply, unless all relevant information as appropriate
is already made available by the instructions for use.
8.4. The label must bear the following particulars which may take the form
of symbols as appropriate:
(a) the name or trade name and address of the manufacturer. For devices imported
into the Community with a view to their distribution in the Community, the
label, the outer packaging, or the instructions for use shall contain in addition
the name and address of the authorized representative of the manufacturer;
(b) the details strictly necessary for the user to uniquely identify the device
and the contents of the packaging;
(c) where appropriate, the word 'STERILE` or a statement indicating any special
microbiological state or state of cleanliness;
(d) the batch code, preceded by the word 'LOT`, or the serial number;
(e) if necessary, an indication of the date by which the device or part of
it should be used, in safety, without degradation of performance, expressed
as the year, the month and, where relevant, the day, in that order;
(f) in case of devices for performance evaluation, the words 'for performance
evaluation only`;
(g) where appropriate, a statement indicating the in vitro use of the device;
(h) any particular storage and/or handling conditions;
(i) where applicable, any particular operating instructions;
(j) appropriate warnings and/or precautions to take;
(k) if the device is intended for self-testing, that fact must be clearly
stated.
8.5. If the intended purpose of the device is not obvious to the user, the
manufacturer must clearly state the intended purpose in the instructions for
use and, if appropriate, on the label.
8.6. Wherever reasonable and practicable, the devices and separate components
must be identified, where appropriate in terms of batches, to allow all appropriate
action to detect any potential risk posed by the devices and detachable components.
8.7. Where appropriate, the instructions for use must contain the following
particulars:
(a) the details referred to in section 8.4 with the exception of points (d)
and (e);
(b) composition of the reagent product by nature and amount or concentration
of the active ingredient(s) of the reagent(s) or kit as well as a statement,
where appropriate, that the device contains other ingredients which might
influence the measurement;
(c) the storage conditions and shelf life following the first opening of the
primary container, together with the storage conditions and stability of working
reagents;
(d) the performances referred to in section 3 of part A;
(e) an indication of any special equipment required including information
necessary for the identification of that special equipment for proper use;
(f) the type of specimen to be used, any special conditions of collection,
pre-treatment and, if necessary, storage conditions and instructions for the
preparation of the patient;
(g) a detailed description of the procedure to be followed in using the device;
(h) the measurement procedure to be followed with the device including as
appropriate: - the principle of the method, - the specific analytical performance
characteristics (e.g. sensitivity, specificity, accuracy, repeatability, reproducibility,
limits of detection and measurement range, including information needed for
the control of known relevant interferences), limitations of the method and
information about the use of available reference measurement procedures and
materials by the user, - the details of any further procedure or handling
needed before the device can be used (for example, reconstitution, incubation,
dilution, instrument checks, etc.), - the indication whether any particular
training is required;
(i) the mathematical approach upon which the calculation of the analytical
result is made;
(j) measures to be taken in the event of changes in the analytical performance
of the device;
(k) information appropriate to users on: - internal quality control including
specific validation procedures, - the traceability of the calibration of the
device;
(l) the reference intervals for the quantities being determined, including
a description of the appropriate reference population;
(m) if the device must be used in combination with or installed with or connected
to other medical devices or equipment in order to operate as required for
its intended purpose, sufficient details of its characteristics to identify
the correct devices or equipment to use in order to obtain a safe and proper
combination;
(n) all the information needed to verify whether the device is properly installed
and can operate correctly and safely, plus details of the nature and frequency
of the maintenance and calibration needed to ensure that the device operates
properly and safely; information about safe waste disposal;
(o) details of any further treatment or handling needed before the device
can be used (for example, sterilization, final assembly, etc.);
(p) the necessary instructions in the event of damage to the protective packaging
and details of appropriate methods of resterilization or decontamination;
(q) if the device is reusable, information on the appropriate processes to
allow reuse, including cleaning, disinfection, packaging and resterilization
or decontamination, and any restriction on the number of reuses;
(r) precautions to be taken as regards exposure, in reasonably foreseeable
environmental conditions, to magnetic fields, external electrical influences,
electrostatic discharge, pressure or variations in pressure, acceleration,
thermal ignition sources, etc.;
(s) precautions to be taken against any special, unusual risks related to
the use or disposal of the device including special protective measures; where
the device includes substances of human or animal origin, attention must be
drawn to their potential infectious nature;
(t) specifications for devices for self-testing: - the results need to be
expressed and presented in a way that is readily understood by a lay person;
information needs to be provided with advice to the user on action to be taken