COUNCIL DIRECTIVE of
20 June 1990 on the approximation of the laws of the Member States relating
to active implantable medical devices (90/385/EEC)
THE COUNCIL OF THE
EUROPEAN COMMUNITIES,
Having regard to the Treaty establishing the European Economic Community,
and in particular Article 100a thereof,
Having regard to the proposal from the Commission (1),
In cooperation with the European Parliament (2),
Having regard to the opinion of the Economic and Social Committee (3),
Whereas in each Member State active implantable medical devices must give
patients, users and other persons a high level of protection and achieve the
intended level of performance when implanted in human beings;
Whereas several Member States have sought to ensure that level of safety by
mandatory specifications relating both to the technical safety features and
the inspection procedures for such devices; whereas those specifications differ
from one Member State to another;
Whereas national provisions ensuring that safety level should be harmonized
in order to guarantee the free movement of active implantable medical devices
without lowering existing and justified levels of safety in the Member States;
Whereas harmonized measures must be distinguished from measures taken by Member
States to manage the financing of public health and sickness insurance schemes
directly or indirectly concerning such devices; whereas, therefore, such provisions
do not affect the right of Member States to implement the abovementioned measures
in compliance with Community law;
Whereas maintaining or improving the level of protection achieved in Member
States constitutes one of this Directive's essential objectives as defined
by the essential requirements;
Whereas rules governing active implantable medical devices can be confined
to those provisions needed to satisfy the OJ No C 149, 18. 6. 1990.
essential requirements; whereas, because they are essential, these requirements
must replace corresponding national provisions;
Whereas, in order to facilitate proof of conformity with these essential requirements
and to permit monitoring of that conformity, it is desirable to have Europe-wide
harmonized standards in respect of the prevention of risks in connection with
the design, manufacture and packaging of active implantable medical devices;
whereas such standards harmonized at European level are drawn up by private-law
bodies and must retain their status as non-mandatory texts; whereas, to that
end, the European Committee for Standardization (CEN) and the European Committee
for Electrotechnical Standardization (Cenelec) are recognized as being the
competent bodies to adopt harmonized standards in accordance with the general
guidelines for cooperation between the Commission and these two bodies, signed
on 13 November 1984; whereas, for the purposes of this Directive, a harmonized
standard is a technical specification (European standard or harmonization
document) adopted by either or both of these bodies, as instructed by the
Commission pursuant to the provisions of Council Directive 83/189/EEC of 28
March 1983 laying down a procedure for the provision of information in the
field of technical standards and regulations (4), as last amended by Directive
88/182/EEC (5), and under the abovementioned general guidelines;
Whereas evaluation procedures have to be established and accepted by common
accord between the Member States in accordance with Community criteria;
Whereas the specific nature of the medical sector makes it advisable to make
provision for the notified body and the manufacturer or his agent established
in the Community to fix, by common accord, the time limits for completion
of the evaluation and verification operations for the conformity of devices,
HAS ADOPTED THIS DIRECTIVE:
Article 1
1. This Directive shall apply to active implantable medical devices.
2. For the purposes of this Directive, the following definitions shall apply:
(a) 'medical device' means any instrument, apparatus, appliance, material
or other article, whether used alone or in combination, together with any
accessories or software for its proper functioning, intended by the manufacturer
to be used for human beings in the: - diagnosis, prevention, monitoring, treatment
or alleviation of disease or injury, - investigation, replacement or modification
of the anatomy or of a physiological process, - control of conception, and
which does not achieve its principal intended action by pharmacological, chemical,
immunological or metabolic means, but which may be assisted in its function
by such means;
(b) 'active medical device' means any medical device relying for its functioning
on a source of electrical energy or any source of power other than that directly
generated by the human body or gravity;
(c) 'active implantable medical device' means any active medical device which
is intended to be totally or partially introduced, surgically or medically,
into the human body or by medical intervention into a natural orifice, and
which is intended to remain after the procedure;
(d) 'custom-made device' means any active implantable medical device specifically
made in accordance with a medical specialist's written prescription which
gives, under his responsibility, specific design characteristics and is intended
to be used only for an individual named patient;
(e) 'device intended for clinical investigation' means any active implantable
medical device intended for use by a specialist doctor when conducting investigations
in an adequate human clinical environment;
(f) 'intended purpose' means the use for which the medical device is intended
and for which it is suited according to the data supplied by the manufacturer
in the instructions;
(g) 'putting into service' means making available to the medical profession
for implantation.
3. Where an active implantable medical device is intended to administer a
substance defined as a medicinal product within the meaning of Council Directive
65/65/EEC of 26 January 1965 on the approximation of provisions laid down
by law, regulation or administrative action relating to proprietary medicinal
products (6), as last amended by Directive 87/21/EEC (7), that substance shall
be subject to the system of marketing authorization provided for in that Directive.
4. Where an active implantable medical device incorporates, as an integral
part, a substance which, if used separately, may be considered to be a medicinal
product within the meaning of Article 1 of Directive 65/65/EEC, that device
must be evaluated and authorized in accordance with the provisions of this
Directive.
5. This Directive constitutes a specific Directive within the meaning of Article
2 (2) of Council Directive 89/336/EEC of 3 May 1989 on the approximation of
the laws of the Member States relating to electromagnetic compatibility (8).
Article 2
Member States shall take all necessary steps to ensure that the devices referred
to in Article 1 (2) (c) and (d) may be placed on the market and put into service
only if they do not compromise the safety and health of patients, users and,
where applicable, other persons when properly implanted, maintained and used
in accordance with their intended purposes.
Article 3
The active implantable medical devices referred to in Article 1 (2) (c), (d)
and (e), hereinafter referred to as 'devices', must satisfy the essential
requirements set out in Annex 1, which shall apply to them account being taken
of the intended purpose of the devices concerned.
Article 4
1. Member States shall not impede the placing on the market or the putting
into service within their territory of devices bearing the CE mark.
2. Member States shall not create any obstacles to: - devices intended for
clinical investigations being made available to specialist doctors for that
purpose if they satisfy the conditions laid down in Article 10 and in Annex
6, - custom-made devices being placed on the market and put into service if
they satisfy the conditions laid down in Annex 6 and are accompanied by the
statement referred to in that Annex. These devices shall not bear the CE mark.
3. At trade fairs, exhibitions, demonstrations, etc., Member States shall
not prevent the showing of devices which do not conform to this Directive,
provided that a visible sign clearly indicates that such devices do not conform
and cannot be put into service until they have been made to comply by the
manufacturer or his authorized representative established within the Community.
4. When a device is put into service, Member States may require the information
described in sections 13, 14 and 15 of Annex 1 to be in their national language(s).
Article 5
Member States shall presume compliance with the essential requirements referred
to in Article 3 in respect of devices which are in conformity with the relevant
national standards adopted pursuant to the harmonized standards the references
of which have been published in the Official Journal of the European Communities;
Member States shall publish the references of such national standards.
Article 6
1. Where a Member State or the Commission considers that the harmonized standards
referred to in Article 5 do not entirely meet the essential requirements referred
to in Article 3, the Commission or the Member State concerned shall bring
the matter before the Standing Committee set up under Directive 83/189/EEC,
giving the reasons therefor. The Committee shall deliver an opinion without
delay. In the light of the opinion of the Committee, the Commission shall
inform Member States of the measures to be taken with regard to the standards
and the publication referred to in Article 5.
2. A Standing Committee, hereinafter referred to as the 'Committee', shall
be set up, composed of the representatives of the Member States and chaired
by the representative of the Commission. The Committee shall draw up its rules
of procedure. Any matter relating to the implementation and practical application
of this Directive may be brought before the Committee, in accordance with
the procedure set out below. The representative of the Commission shall submit
to the Committee a draft of the measures to be taken. The Committee shall
deliver its opinion according to the urgency of the matter, if necessary by
taking a vote. The opinion shall be recorded in the minutes; in addition,
each Member State shall have the right to ask to have its position recorded
in the minutes. The Commission shall take the utmost account of the opinion
delivered by the Committee. It shall inform the Committee of the manner in
which its opinion has been taken into account.
Article 7
1. Where a Member State finds that the devices referred to in Article 1 (2)
(c) and (d), correctly put into service and used in accordance with their
intended purpose, may compromise the health and/or safety of patients, users
or, where applicable, other persons, it shall take all appropriate measures
to withdraw such devices from the market or prohibit or restrict their being
placed on the market or their being put into service. The Member State shall
immediately inform the Commission of any such measure, indicating the reasons
for its decision and, in particular, whether non-compliance with this Directive
is due to:
(a) failure to meet the essential requirements referred to in Article 3, where
the device does not meet in full or in part the standards referred to in Article
5;
(b) incorrect application of those standards; (c) shortcomings in the standards
themselves.
2. The Commission shall enter into consultation with the parties concerned
as soon as possible. Where, after such consultation, the Commission finds
that: - the measures are justified, it shall immediately so inform the Member
State which took the initiative and the other Member States; where the decision
referred to in paragraph 1 is attributed to shortcomings in the standards,
the Commission shall, after consulting the parties concerned, bring the matter
before the Committee referred to in Article 6 (1) within two months if the
Member State which has taken the decision intends to maintain it and shall
initiate the procedures referred to in Article 6 (1), - the measures are unjustified,
it shall immediately so inform the Member State which took the initiative
and the manufacturer or his authorized representative established within the
Community.
3. Where a device which does not comply bears the CE mark, the competent Member
State shall take appropriate action against whomsoever has affixed the mark
and shall inform the Commission and the other Member States thereof.
4. The Commission shall ensure that the Member States are kept informed of
the progress and outcome of this procedure.
Article 8
1. Member States shall take the necessary steps to ensure that information
brought to their knowledge regarding the incidents mentioned below involving
a device is recorded and evaluated in a centralized manner:
(a) any deterioration in the characteristics and performances of a device,
as well as any inaccuracies in the instruction leaflet which might lead to
or might have led to the death of a patient or to a deterioration in his state
of health;
(b) any technical or medical reason resulting in withdrawal of a device from
the market by the manufacturer.
2. Member States shall, without prejudice to Article 7, forthwith inform the
Commission and the other Member States of the incidents referred to in paragraph
1 and of the relevant measures taken or contemplated.
Article 9
1. In the case of devices other than those which are custom-made or intended
for clinical investigations, the manufacturer must, in order to affix the
CE mark, at his own choice:
(a) follow the procedure relating to the EC declaration of conformity set
out in Annex 2; or
(b) follow the procedure relating to EC type-examination set out in Annex
3, coupled with: i(i) the procedure relating to EC verification set out in
Annex 4, or (ii) the procedure relating to the EC declaration of conformity
to type set out in Annex 5.
2. In the case of custom-made devices, the manufacturer must draw up the declaration
provided for in Annex 6 before placing each device on the market.
3. Where appropriate, the procedures provided for in Annexes 3, 4 and 6 may
be discharged by the manufacturer's authorized representative established
in the Community.
4. The records and correspondence relating to the procedures referred to in
paragraphs 1, 2 and 3 shall be in an official language of the Member State
in which the said procedures will be carried out and/or in a language acceptable
to the notified body defined in Article 11.
Article 10
1. In the case of devices intended for clinical investigations, the manufacturer
or his authorized representative established in the Community shall, at least
60 days before the commencement of the investigations, submit the statement
referred to in Annex 6 to the competent authorities of the Member State in
which the investigations are to be conducted.
2. The manufacturer may commence the relevant clinical investigations at the
end of a period of 60 days after notification, unless the competent authorities
have notified him within that period of a decision to the contrary, based
on considerations of public health or public order.
3. The Member States shall, if necessary, take the appropriate steps to ensure
public health and order.
Article 11
1. Each Member State shall notify the other Member States and the Commission
of the bodies which they have designated for carrying out the tasks pertaining
to the procedures referred to in Articles 9 and 13, the specific tasks for
which each body has been designated and the identifying logo of these bodies,
hereinafter referred to as 'notified bodies'. The Commission shall publish
a list of these notified bodies, together with the tasks for which they have
been notified, in the Official Journal of the European Communities and shall
ensure that the list is kept up to date.
2. Member States shall apply the minimum criteria, set out in Annex 8, for
the designation of bodies. Bodies that satisfy the criteria fixed by the relevant
harmonized standards shall be presumed to satisfy the relevant minimum criteria.
3. A Member State that has notified a body shall withdraw that notification
if it finds that the body no longer meets the criteria referred to in paragraph
2. It shall immediately inform the other Member States and the Commission
thereof.
4. The notified body and the manufacturer or his agent established in the
Community shall fix, by common accord, the time limits for completion of the
evaluation and verification operations referred to in Annexes 2 to 5.
Article 12
1. Devices other than those which are custom made or intended for clinical
investigations considered to meet the essential requirements referred to in
Article 3 must bear the EC mark of conformity.
2. The EC mark of conformity, as shown in Annex 9, must appear in a visible,
legible and indelible form on the sterile pack and, where appropriate, on
the sales packaging, if any, and on the instruction leaflet. It must be accompanied
by the logo of the notified body responsible for implementation of the procedures
set out in Annexes 2, 4 and 5.
3. The affixing of marks likely to be confused with the EC mark of conformity
shall be prohibited.
Article 13
Where it is established that the EC mark has been wrongly affixed, in particular,
in respect of devices: - that do not conform to the relevant standards referred
to in Article 5, should the manufacturer have opted for conformity therewith,
- that do not conform to an approved type, - that conform to an approved type
which does not meet the relevant essential requirements, - regarding which
the manufacturer has failed to fulfil his obligations under the relevant EC
declaration of conformity, the notified body shall take appropriate measures
and forthwith inform the competent Member State thereof.
Article 14
Any decision taken pursuant to this Directive and resulting in the refusal
of or restrictions on the placing on the market and/or putting into service
of a device shall state the exact grounds on which it is based. Such decision
shall be notified without delay to the party concerned, who shall at the same
time be informed of the remedies available to him under the laws in force
in the Member State in question and of the time limits to which such remedies
are subject.
Article 15
Member States shall ensure that all the parties involved in the application
of this Directive are bound to observe confidentiality with regard to all
information obtained in carrying out their tasks. This does not affect the
obligations of Member States and notified bodies with regard to mutual information
and the dissemination of warnings.
Article 16
1. Before 1 July 1992, Member States shall adopt and publish the laws,
regulations and administrative provisions necessary in order to comply with
this Directive. They shall forthwith inform the Commission thereof. They shall
apply such provisions from 1 January 1993.
2. Member States shall communicate to the Commission the texts of the provisions
of national law which they adopt in the field covered by this Directive.
3. Member States shall, for the period up to 31 December 1994, permit the
placing on the market and putting into service of devices complying with national
rules in force in their territory on 31 December 1992. Article 17 This Directive
is addressed to the Member States.
ANNEX 1
ESSENTIAL REQUIREMENTS
I. GENERAL REQUIREMENTS
1. The devices must be designed and manufactured in such a way that, when
implanted under the conditions and for the purposes laid down, their use does
not compromise the clinical condition or the safety of patients. They must
not present any risk to the persons implanting them or, where applicable,
to other persons.
2. The devices must achieve the performances intended by the manufacturer,
viz. be designed and manufactured in such a way that they are suitable for
one or more of the functions referred to in Article 1 (2) (a) as specified
by him.
3. The characteristics and performances referred to in sections 1 and 2 must
not be adversely affected to such a degree that the clinical condition and
safety of the patients or, as appropriate, of other persons are compromised
during the lifetime of the device anticipated by the manufacturer, where the
device is subjected to stresses which may occur during normal conditions of
use.
4. The devices must be designed, manufactured and packed in such a way that
their characteristics and performances are not adversely affected in the storage
and transport conditions laid down by the manufacturer (temperature, humidity,
etc.).
5. Any side effects or undesirable conditions must constitute acceptable risks
when weighed against the performances intended.
II. REQUIREMENTS REGARDING DESIGN AND CONSTRUCTION
6. The solutions adopted by the manufacturer for the design and construction
of the devices must comply with safety principles taking account of the generally
acknowledged state of the art. 7. Implantable devices must be designed, manufactured
and packed in a non-reusable pack according to appropriate procedures to ensure
they are sterile when placed on the market and, in the storage and transport
conditions stipulated by the manufacturer, remain so until the packaging is
removed and they are implanted.
8. Devices must be designed and manufactured in such a way as to remove or
minimize as far as possible: - the risk of physical injury in connection with
their physical, including dimensional, features, - risks connected with the
use of energy sources with particular reference, where electricity is used,
to insulation, leakage currents and overheating of the devices, - risks connected
with reasonably foreseeable environmental conditions such as magnetic fields,
external electrical influences, electrostatic discharge, pressure or variations
in pressure and acceleration, - risks connected with medical treatment, in
particular those resulting from the use of defibrillators or high-frequency
surgical equipment, - risks connected with ionizing radiation from radioactive
substances included in the device, in compliance with the protection requirements
laid down in Directive 80/836/Euratom (;), as amended by Directives 84/467/Euratom
($) and 84/466/Euratom (=), - risks which may arise where maintenance and
calibration are impossible, including: - excessive increase of leakage currents,
- aging of the materials used, - excess heat generated by the device, - decreased
accuracy of any measuring or control mechanism.
(;) OJ No L 246, 17. 9. 1980, p. 1.
($) OJ No L 265, 5. 10. 1984, p. 4.
(=) OJ No L 265, 5. 10. 1984, p. 1.
9. The devices must be designed and manufactured in such a way as to guarantee
the characteristics and performances referred to in I. 'General requirements',
with particular attention being paid to: - the choice of materials used, particularly
as regards toxicity aspects, - mutual compatibility between the materials
used and biological tissues, cells and body fluids, account being taken of
the anticipated use of the device, - compatibility of the devices with the
substances they are intended to administer, - the quality of the connections,
particularly in respect of safety, - the reliability of the source of energy,
- if appropriate, that they are leakproof, - proper functioning of the programming
and control systems, including software.
10. Where a device incorporates, as an integral part, a substance which, when
used separately, is likely to be considered to be a medicinal product as defined
in Article 1 of Directive 65/65/EEC, and whose action in combination with
the device may result in its bioavailability, the safety, quality and usefulness
of the substance, account being taken of the purpose of the device, must be
verified by analogy with the appropriate methods specified in Directive 75/318/EEC
(;), as last amended by Directive 89/341/EEC ($).
11. The devices and, if appropriate, their component parts must be identified
to allow any necessary measure to be taken following the discovery of a potential
risk in connection with the devices and their component parts.
12. Devices must bear a code by which they and their manufacturer can be unequivocally
identified (particularly with regard to the type of device and year of manufacture);
it must be possible to read this code, if necessary, without the need for
a surgical operation.
13. When a device or its accessories bear instructions required for the operation
of the device or indicate operating or adjustment parameters, by means of
a visual system, such information must be understandable to the user and,
as appropriate, the patient.
14. Every device must bear, legibly and indelibly, the following particulars,
where appropriate in the form of generally recognized symbols:
14.1. On the sterile pack: - the method of sterilization, - an indication
permitting this packaging to be recognized as such, - the name and address
of the manufacturer, - a description of the device, - if the device is intended
for clinical investigations, the words: 'exclusively for clinical investigations',
- if the device is custom-made, the words 'custom-made device', - a declaration
that the implantable device is in a sterile condition, - the month and year
of manufacture, - an indication of the time limit for implanting a device
safely.
14.2. On the sales packaging: - the name and address of the manufacturer,
- a description of the device, - the purpose of the device, - the relevant
characteristics for its use, - if the device is intended for clinical investigations,
the words: 'exclusively for clinical investigations',
(;) OJ No L 147, 9. 6. 1975, p. 1.
($) OJ No L 142, 25. 5. 1989, p. 11. - if the device is custom-made, the words:
'custom-made device', - a declaration that the implantable device is in a
sterile condition, - the month and year of manufacture, - an indication of
the time limit for implanting a device safely, - the conditions for transporting
and storing the device.
15. When placed on the market, each device must be accompanied by instructions
for use giving the following particulars: - the year of authorization to affix
the CE mark, - the details referred to in 14.1 and 14.2, with the exception
of those referred to in the eighth and ninth indents, - the performances referred
to in section 2 and any undesirable side effects, - information allowing the
physician to select a suitable device and the corresponding software and accessories,
- information constituting the instructions for use allowing the physician
and, where appropriate, the patient to use the device, its accessories and
software correctly, as well as information on the nature, scope and times
for operating controls and trials and, where appropriate, maintenance measures,
- information allowing, if appropriate, certain risks in connection with implantation
of the device to be avoided, - information regarding the risks of reciprocal
interference (*) in connection with the presence of the device during specific
investigations or treatment, - the necessary instructions in the event of
the sterile pack being damaged and, where appropriate, details of appropriate
methods of resterilization, - an indication, if appropriate, that a device
can be reused only if it is reconditioned under the responsibility of the
manufacturer to comply with the essential requirements. The instruction leaflet
must also include details allowing the physician to brief the patient on the
contra-indications and the precautions to be taken. These details should cover
in particular: - information allowing the lifetime of the energy source to
be established, - precautions to be taken should changes occur in the device's
performance, - precautions to be taken as regards exposure, in reasonably
foreseeable environmental conditions, to magnetic fields, external electrical
influences, electrostatic discharge, pressure or variations in pressure, acceleration,
etc., - adequate information regarding the medicinal products which the device
in question is designed to administer.
16. Confirmation that the device satisfies the requirements in respect of
characteristics and performances, as referred to in I. 'General requirements',
in normal conditions of use, and the evaluation of the side effects or undesirable
effects must be based on clinical data established in accordance with Annex
7. (*) 'Risks of reciprocal interference' means adverse effects on the device
caused by instruments present at the time of investigations or treatment,
and vice versa.
ANNEX 2
EC DECLARATION OF CONFORMITY (Complete quality assurance system)
1. The manufacturer shall apply the quality system approved for the design,
manufacture and final inspection of the products concerned as specified in
sections 3 and 4 and shall be subject to EC surveillance as specified in section
5.
2. The declaration of conformity is the procedure by means of which the manufacturer
who satisfies the obligations of section 1 ensures and declares that the products
concerned meet the provisions of this Directive which apply to them The manufacturer
shall apply the CE mark in accordance with Article 12 and draw up a written
declaration of conformity. This declaration shall cover one or more identified
specimens of the product and shall be kept by the manufacturer. The CE mark
shall be accompanied by the identifying logo of the notified body responsible.
3. Quality system
3.1. The manufacturer shall make an application for evaluation of his quality
system to a notified body. The application shall include: - all the appropriate
items of information for the category of products manufacture of which is
envisaged, - the quality-system documentation, - an undertaking to fulfil
the obligations arising from the quality system as approved, - an undertaking
to maintain the approved quality system in such a way that it remains adequate
and efficacious, - an undertaking by the manufacturer to institute and keep
up-dated a post-marketing surveillance system. The undertaking shall include
an obligation for the manufacturer to notify the competent authorities of
the following incidents immediately on learning of them: i(i) any deterioration
in the characteristics or performances, and any inaccuracies in the instruction
leaflet for a device which might lead to or have led to the death of a patient
or a deterioration in his state of health; (ii) any technical or medical reason
resulting in withdrawal of a device from the market by the manufacturer.
3.2. The application of the quality system must ensure that the products conform
to the provisions of this Directive which apply to them at every stage, from
design to final controls. All the elements, requirements and provisions adopted
by the manufacturer for his quality system shall be documented in a systematic
and orderly manner in the form of written policies and procedures. This quality-system
documentation must make possible a uniform interpretation of the quality policies
and procedures such as quality programs, quality plans, quality manuals
and quality records. It shall include in particular an adequate description
of:
(a) the manufacturer's quality objectives;
(b) the organization of the business and in particular: - the organizational
structures, the responsibilities of the managerial staff and their organizational
authority where quality of design and manufacture of the products is concerned,
- the methods of monitoring the efficient operation of the quality system
and in particular its ability to achieve the desired quality of the design
and of the products, including control of products which do not conform;
(c) the procedures for monitoring and verifying the design of the products
and in particular: - the design specifications, including the standards which
will be applied and a description of the solutions adopted to fulfil the essential
requirements which apply to the products when the standards referred to in
Article 5 are not applied in full, - the techniques of control and verification
of the design, the processes and systematic actions which will be used when
the products are being designed;
(d) the techniques of control and of quality assurance at the manufacturing
stage and in particular: - the processes and procedures which will be used,
particularly as regards sterilization, purchasing and the relevant documents,
- product-identification procedures drawn up and kept up to date from drawings,
specifications or other relevant documents at every stage of manufacture;
(e) the appropriate tests and trials which will be effected before, during
and after production, the frequency with which they will take place, and the
test equipment used. 3.3. Without prejudice to Article 13 of this Directive,
the notified body shall effect an audit of the quality system to determine
whether it meets the requirements referred to in 3.2. It shall presume conformity
with these requirements for the quality systems which use the corresponding
harmonized standards. The team entrusted with the evaluation shall include
at least one member who has already had experience of evaluations of the technology
concerned. The evaluation procedure shall include an inspection on the manufacturer's
premises. The decision shall be notified to the manufacturer after the final
inspection. It shall contain the conclusions of the control and a reasoned
evaluation. 3.4. The manufacturer shall inform the notified body which has
approved the quality system of any plan to alter the quality system. The notified
body shall evaluate the proposed modifications and shall verify whether the
quality system so modified would meet the requirements referred to in 3.2;
it shall notify the manufacturer of its decision. This decision shall contain
the conclusions of the control and a reasoned evaluation.
4. Examination of the design of the product
4.1. In addition to the obligations incumbent on him under section 3, the
manufacturer shall make an application for examination of the design dossier
relating to the product which he plans to manufacture and which falls into
the category referred to in 3.1.
4.2. The application shall describe the design, manufacture, and performances
of the product in question and shall include the necessary particulars which
make it possible to evaluate whether it complies with the requirements of
this Directive. It shall include inter alia: - the design specifications,
including the standards which have been applied, - the necessary proof of
their appropriations, in particular where the standards referred to in Article
5 have not been applied in full. This proof must include the results of the
appropriate tests carried out by the manufacturer or carried out under his
responsibility, - a statement as to whether or not the device incorporates,
as an integral part, a substance as referred to in section 10 of Annex 1,
whose action in combination with the device may result in its bioavailability,
together with data on the relevant trials conducted, - the clinical data referred
to in Annex 7, - the draft instruction leaflet.
4.3. The notified body shall examine the application and, where the product
complies with the relevant provisions of this Directive, shall issue the applicant
with an EC design examination certificate. The notified body may require the
application to be supplemented by further tests or proof so that compliance
with the requirements of the Directive may be evaluated. The certificate shall
contain the conclusions of the examination, the conditions of its validity,
the data needed for identification of the approved design and, where appropriate,
a description of the intended use of the product.
4.4. The applicant shall inform the notified body which issued the EC design
examination certificate of any modification made to the approved design. Modifications
made to the approved design must obtain supplementary approval from the notified
body which issued the EC design examination certificate where such modifications
may affect conformity with the essential requirements of this Directive or
the conditions prescribed for the use of the product. This supplementary approval
shall be given in the form of an addendum to the EC design examination certificate.
5. Surveillance
5.1. The aim of surveillance is to ensure that the manufacturer duly fulfils
the obligations arising from the approved quality system.
5.2. The manufacturer shall authorize the notified body to carry out all necessary
inspections and shall supply it with all appropriate information, in particular:
- the quality-system documentation, - the data stipulated in the part of the
quality system relating to design, such as the results of analyses, calculations,
tests, etc., - the data stipulated in the part of the quality system relating
to manufacture, such as reports concerning inspections, tests, standardizations/calibrations
and the qualifications of the staff concerned, etc.
5.3. The notified body shall periodically carry out appropriate inspections
and evaluations in order to ascertain that the manufacturer is applying the
approved quality system, and shall supply the manufacturer with an evaluation
report.
5.4. In addition, the notified body may make unannounced visits to the manufacturer,
and shall supply him with an inspection report. 6. The notified body shall
communicate to the other notified bodies all relevant information concerning
approvals of quality systems issued, refused and withdrawn.
ANNEX 3
EC TYPE-EXAMINATION
1. EC type-examination is the procedure whereby a notified body observes and
certifies that a representative sample of the production envisaged satisfies
the relevant provisions of this Directive.
2. The application for EC type-examination shall be made by the manufacturer,
or by his authorized representative established in the Community, to a notified
body. The application shall include: - the name and address of the manufacturer
and the name and address of the authorized representative if the application
is made by the latter, - a written declaration specifying that an application
has not been made to any other notified body, - the documentation described
in section 3 needed to allow an evaluation to be made of the conformity of
a representative sample of the production in question, hereinafter referred
to as 'type', with the requirements of this Directive. The applicant shall
make a 'type' available to the notified body. The notified body may request
other samples as necessary.
3. The documentation must make it possible to understand the design, the manufacture
and the performances of the product. The documentation shall contain the following
items in particular: - a general description of the type, - design drawings,
methods of manufacture envisaged, in particular as regards sterilization,
and diagrams of parts, sub-assemblies, circuits, etc., - the descriptions
and explanations necessary for the understanding of the abovementioned drawings
and diagrams and of the operation of the product, - a list of the standards
referred to in Article 5, applied in full or in part, and a description of
the solutions adopted to satisfy the essential requirements where the standards
referred to in Article 5 have not been applied, - the results of design calculations,
investigations and technical tests carried out, etc., - a statement as to
whether or not the device incorporates, as an integral part, a substance as
referred to in section 10 of Annex 1, whose action in combination with the
device may result in its bioavailability, together with data on the relevant
trials conducted, - the clinical data referred to in Annex 7, - the draft
instruction leaflet.
4. The notified body shall:
4.1. examine and evaluate the documentation, verify that the type has been
manufactured in accordance with that documentation; it shall also record the
items which have been designed in accordance with the applicable provisions
of the standards referred to in Article 5, as well as the items for which
the design is not based on the relevant provisions of the said standards;
4.2. carry out or have carried out the appropriate inspections and the tests
necessary to verify whether the solutions adopted by the manufacturer satisfy
the essential requirements of this Directive where the standards referred
to in Article 5 have not been applied;
4.3. carry out or have carried out the appropriate inspections and the tests
necessary to verify whether, where the manufacturer has chosen to apply the
relevant standards, these have actually been applied;
4.4. agree with the applicant on the place where the necessary inspections
and tests will be carried out.
5. Where the type meets the provisions of this Directive, the notified body
shall issue an EC type-examination certificate to the applicant. The certificate
shall contain the name and address of the manufacturer, the conclusions of
the control, the conditions under which the certificate is valid and the information
necessary for identification of the type approved. The significant parts of
the documentation shall be attached to the certificate and a copy shall be
kept by the notified body.
6. The applicant shall inform the notified body which issued the EC type-examination
certificate of any modification made to the approved product. Modifications
to the approved product must receive further approval from the notified body
which issued the EC type-examination certificate where such modifications
may affect conformity with the essential requirements or with the conditions
of use specified for the product. This new approval shall be issued, where
appropriate, in the form of a supplement to the initial EC type-examination
certificate.
7. Each notified body shall communicate to the other notified bodies all relevant
information on EC-type examination certificates and supplements issued, refused
or withdrawn.
8. Other notified bodies may obtain a copy of the EC type-examination certificates
and/or the supplements to them. The annexes to the certificates shall be made
available to the other notified bodies when a reasoned application is made
and after first informing the manufacturer.
ANNEX 4
EC VERIFICATION
1. EC verification is the act by which a notified body verifies and certifies
that products conform to the type described in the EC type-examination certificate
and satisfy the relevant requirements of this Directive.
2. The manufacturer shall, before the start of manufacture, prepare documents
defining the manufacturing process, in particular as regards sterilization,
together with all the routine, pre-established provisions to be implemented
to ensure homogeneity of production and conformity of the products with the
type described in the EC type-examination certificate as well as with the
relevant requirements of the Directive.
3. The manufacturer shall undertake to institute and keep up-dated a post-marketing
surveillance system. The undertaking shall include an obligation for the manufacturer
to notify the competent authorities of the following events immediately on
learning of them: i(i) any deterioration in the characteristics or performances,
and any inaccuracies in the instruction leaflet for a device which might lead
to or have led to the death of a patient or a deterioration in his state of
health; (ii) any technical or medical reason resulting in withdrawal of a
device from the market by the manufacturer.
4. The notified body shall carry out EC verification by controls and tests
on the products on a statistical basis as specified in 5. The manufacturer
must authorize the notified body to evaluate the efficiency of the measures
taken pursuant to section 2, by audit where appropriate.
5. Statistical verification
5.1. The manufacturer shall present the manufactured products in the form
of homogeneous batches.
5.2. A random sample shall be taken from each batch. The products which make
up the sample shall be examined individually and appropriate tests, defined
in the relevant standard(s) referred to in Article 5, or equivalent tests,
shall be carried out to verify the conformity of the products with the type
described in the EC type-examination certificate, in order to determine whether
the batch is to be accepted or rejected.
5.3. Statistical control of products will be based on attributes, entailing
a sampling system with the following characteristics: - a level of quality
corresponding to a probability of acceptance of 95 %, with a non-conformity
percentage of between 0,29 and 1 %, - a limit quality corresponding to a probability
of acceptance of 5 %, with a non-conformity percentage of between 3 and 7
%.
5.4. If a batch is accepted, the notified body shall draw up a written certificate
of conformity. All the products in the batch may be placed on the market,
with the exception of those products in the sample which were found not to
conform. If a batch is rejected, the notified body which is responsible shall
take the appropriate measures to prevent the batch from being placed on the
market. If justified on practical grounds, the manufacturer may affix the
CE mark during manufacture, under the responsibility of the notified body,
in accordance with Article 12, accompanied by the identifying logo of the
notified body responsible for statistical verification.
ANNEX 5
EC DECLARATION OF CONFORMITY TO TYPE (Assurance of production quality)
1. The manufacturer shall apply the quality system approved for the manufacture
and shall conduct the final inspection of the products concerned as specified
in 3; he shall be subject to the surveillance referred to in section 4.
2. This declaration of conformity is the procedural element whereby the manufacturer
who satisfies the obligations of section 1 guarantees and declares that the
products concerned conform to the type described in the EC type-examination
certificate and meet the provisions of the Directive which apply to them.
The manufacturer shall affix the CE mark in accordance with Article 12 and
draw up a written declaration of conformity. This declaration shall cover
one or more identified specimens of the product and shall be kept by the manufacturer.
The CE mark shall be accompanied by the identifying logo of the notified body
responsible.
3. Quality system
3.1. The manufacturer shall make an application for evaluation of his quality
system to a notified body. The application shall include: - all appropriate
information concerning the products which it is intended to manufacture, -
the quality-system documentation, - an undertaking to fulfil the obligations
arising from the quality system as approved, - an undertaking to maintain
the approved quality system in such a way that it remains adequate and efficacious,
- where appropriate, the technical documentation relating to the approved
type and a copy of the EC type-examination certificate, - an undertaking by
the manufacturer to institute and keep up-dated a post-marketing surveillance
system. The undertaking shall include an obligation for the manufacturer to
notify the competent authorities of the following incidents immediately on
learning of them: i(i) any deterioration in the characteristics or performances,
and any inaccuracies in the instruction leaflet for a device which might lead
to or have led to the death of a patient or a deterioration in his state of
health; (ii) any technical or medical reason resulting in withdrawal of a
device from the market by the manufacturer.
3.2. Application of the quality system must ensure that the products conform
to the type described in the EC type-examination certificate. All the elements,
requirements and provisions adopted by the manufacturer for his quality system
shall be documented in a systematic and orderly manner in the form of written
policies and procedures. This quality-system documentation must make possible
a uniform interpretation of the quality policies and procedures such as quality
programs, quality plans, quality manuals and quality records. It shall include
in particular an adequate description of:
(a) the manufacturer's quality objectives;
(b) the organization of the business and in particular: - the organizational
structures, the responsibilities of the managerial staff and their organizational
authority where manufacture of the products is concerned, - the methods of
monitoring the efficient operation of the quality system and in particular
its ability to achieve the desired quality of the design and of the products,
including control of products which do not conform;
(c) the techniques of control and of quality assurance at the manufacturing
stage and in particular: - the processes and procedures which will be used,
particularly as regards sterilization, purchasing and the relevant documents,
- product identification procedures drawn up and kept up-to-date from drawings,
specifications or other relevant documents at every stage of manufacture;
(d) the appropriate tests and trials which will be effected before, during
and after production, the frequency with which they will take place, and the
test equipment used.
3.3. Without prejudice to Article 13, the notified body shall effect an audit
of the quality system to determine whether it meets the requirements referred
to in 3.2. It shall presume conformity with these requirements for the quality
systems which use the corresponding harmonized standards. The team entrusted
with the evaluation shall include at least one member who has already had
experience of evaluations of the technology concerned. The evaluation procedure
shall include an inspection on the manufacturer's premises. The decision shall
be notified to the manufacturer after the final inspection. It shall contain
the conclusions of the control and a reasoned evaluation.
3.4. The manufacturer shall inform the notified body which has approved the
quality system of any plan to alter that system. The notified body shall evaluate
the proposed modifications and shall verify whether the quality system so
modified would meet the requirements referred to in 3.2; it shall notify the
manufacturer of its decision. This decision shall contain the conclusions
of the control and a reasoned evaluation.
4. Surveillance
4.1. The aim of surveillance is to ensure that the manufacturer duly fulfils
the obligations which arise from the approved quality system.
4.2. The manufacturer shall authorize the notified body to carry out all necessary
inspections and shall supply it with all appropriate information, in particular:
- the quality-system documentation, - the data stipulated in the part of the
quality system relating to manufacture, such as reports concerning inspections,
tests, standardizations/calibrations and the qualifications of the staff concerned,
etc.
4.3. The notified body shall periodically carry out appropriate inspections
and evaluations in order to ascertain that the manufacturer is applying the
approved quality system, and shall supply the manufacturer with an evaluation
report.
4.4. In addition, the notified body may make unannounced visits to the manufacturer,
and shall supply him with an inspection report.
5. The notified body shall communicate to the other notified bodies all relevant
information concerning approvals of quality systems issued, refused or withdrawn.
ANNEX 6
STATEMENT CONCERNING DEVICES INTENDED FOR SPECIAL PURPOSES
1. The manufacturer or his authorized representative established within the
Community shall draw up for custom-made devices or for devices intended for
clinical investigations the statement comprising the elements stipulated in
section 2.
2. The statement shall comprise the following information:
2.1. For custom-made devices: - data allowing the device in question to be
identified, - a statement affirming that the device is intended for exclusive
use by a particular patient, together with his name, - the name of the doctor
who drew up the prescription and, if applicable, the name of the clinic concerned,
- the particular features of the device as described by the medical prescription
concerned, - a statement affirming that the device complies with the essential
requirements given in Annex 1 and, where applicable, indicating which essential
requirements have not been wholly met, together with the grounds.
2.2. For devices intended for clinical investigations covered in Annex 7:
- data allowing the devices in question to be identified, - an investigation
plan giving in particular the purpose, scope and number of the devices concerned,
- the name of the doctor and of the institution responsible for the investigations,
- the place, date of commencement and duration scheduled for the investigations,
- a statement affirming that the device in question complies with the essential
requirements apart from the aspects constituting the object of the investigations
and that, with regard to these aspects, every precaution has been taken to
protect the health and safety of the patient.
3. The manufacturer shall undertake to keep available for the competent national
authorities:
3.1. For custom-made devices, documentation enabling the design, manufacture
and performances of the product, including the expected performances, to be
understood, so as to allow conformity with the requirement of this Directive
to be assessed. The manufacturer shall take all necessary measures to see
that the manufacturing process ensures that the products manufactured conform
to the documentation referred to in the first paragraph.
3.2. For devices intended for clinical investigations, the documentation shall
also contain: - a general description of the product, - design drawings, manufacturing
methods, in particular as regards sterilization, and diagrams of parts, sub-assemblies,
circuits, etc., - the descriptions and explanations necessary for the understanding
of the said drawings and diagrams and of the operation of the product, - a
list of the standards laid down in Article 5, applied in full or in part,
and a description of the solutions adopted to satisfy the essential requirements
of the Directive where the standards in Article 5 have not been applied, -
the results of the design calculations, checks and technical tests carried
out, etc. The manufacturer shall take all necessary measures to see that the
manufacturing process ensures that the products manufactured conform to the
documentation referred to in 3.1 and in in the first paragraph of this section.
The manufacturer may authorize the evaluation, by audit where necessary, of
the effectiveness of these measures.
ANNEX 7
CLINICAL EVALUATION
1. General provisions
1.1. Adequacy of the clinical data presented, as referred to in section 4.2
of Annex 2, and in section 3 of Annex 3, shall be based, account being taken
as appropriate of the relevant harmonized standards, on either:
1.1.1. a collation of currently available relevant scientific literature covering
the intended use of the device and the techniques therefor, as well as, if
appropriate, a written report making a critical assessment of this collation;
or
1.1.2. the results of all clinical investigations made, including those carried
out in accordance with section 2.
1.2. All data must remain confidential unless it is deemed essential that
they be divulged.
2. Clinical investigation
2.1. Purpose The purpose of clinical investigation is to: - verify that, under
normal conditions of use, the performances of the device comply with those
indicated in section 2 of Annex 1, - determine any undesirable side effects,
under normal conditions of use, and assess whether they are acceptable risks
having regard to the intended performance of the device.
2.2. Ethical consideration Clinical investigations shall be made in accordance
with the Declaration of Helsinki approved by the 18th World Medical Assembly
in Helsinki, Finland, in 1964, and amended by the 29th World Medical Assembly
in Tokyo, Japan, in 1975 and the 35th World Medical Assembly in Venice, Italy,
in 1983. It is mandatory that all measures relating to the protection of human
subjects are carried out in the spirit of the Declaration of Helsinki. This
includes every step in the clinical investigation from first consideration
of need and justification of the study to publication of results.
2.3.
Methods 2.3.1. Clinical investigations shall be performed according to an
appropriate state of the art plan of investigation defined in such a way as
to confirm or refute the manufacturer's claims for the device; the investigations
shall include an adequate number of observations to guarantee the scientific
validity of the conclusions.
2.3.2. The procedures utilized to perform the investigations shall be appropriate
to the device under examination.
2.3.3. Clinical investigations shall be performed in circumstances equivalent
to those which would be found in normal conditions of use of the device.
2.3.4. All appropriate features, including those involving the safety and
performances of the device, and its effects on the patients, shall be examined.
2.3.5. All adverse events shall be fully recorded.
2.3.6. The investigations shall be performed under the responsibility of an
appropriately qualified medical specialist, in an appropriate environment.
The medical specialist shall have access to the technical data regarding the
device.
2.3.7. The written report, signed by the responsible medical specialist, shall
comprise a critical evaluation of all the data collected during the clinical
investigation.
ANNEX 8
MINIMUM CRITERIA TO BE MET WHEN DESIGNATING INSPECTION BODIES TO BE NOTIFIED
1. The body, its director and the staff responsible for carrying out the evaluation
and verification operations shall not be the designer, manufacturer, supplier
or installer of devices which they control, nor the authorized representative
of any of those parties. They may not become directly involved in the design,
construction, marketing or maintenance of the devices, nor represent the parties
engaged in these activities. This does not preclude the possibility of exchanges
of technical information between the manufacturer and the body.
2. The body and its staff must carry out the evaluation and verification operations
with the highest degree of professional integrity and technical competence
and must be free from all pressures and inducements, particularly financial,
which might influence their judgement or the results of the inspection, especially
from persons or groups of persons with an interest in the results of verifications.
3. The body must be able to carry out all the tasks in one of Annexes 2 to
5 assigned to such a body and for which it has been notified, whether those
tasks are carried out by the body itself or under its responsibility. In particular,
it must have at its disposal the necessary staff and possess the necessary
facilities to enable it to perform properly the technical and administrative
tasks connected with evaluation and verification; it must also have access
to the equipment necessary for the verifications required.
4. The staff responsible for control operations must have: - sound vocational
training covering all the evaluation and verification operations for which
the body has been designated, - satisfactory knowledge of the requirements
of the controls they carry out and adequate experience of such operations,
- the ability required to draw up the certificates, records and reports to
demonstrate that the controls have been carried out.
5. The impartiality of inspection staff must be guaranteed. Their remuneration
must not depend on the number of controls carried out, nor on the results
of such controls.
6. The body must take out liability insurance unless liability is assumed
by the State in accordance with national law, or the Member State itself is
directly responsible for controls.
7. The staff of the body are bound to observe professional secrecy with regard
to all information gained in carrying out their tasks (except vis-à-vis the
competent administrative authorities of the State in which their activities
are carried out) under this Directive or any provision of national law giving
effect to it.
ANNEX 9
CE MARK OF CONFORMITY